13.4. TSurge Menu
TSurge test are possible in the menu TSurge Quickstart. This menu is only available if the CTRL cable between
TSurge7 module and UCS500N7 is connected and the TSurge is detected by the UCS500N7.
The menu offers different test routines for ring wave testing.
TSURGE
F1 Quick Start
Easy and fast online-operation of the
equipment.
F1 : TSurge Quick Start
F1
F2
F3
F4
F5
F6
F7
13.4.1.
Quick Start
Easy and very fast operation of all standard functions of the equipment. The latest simulator settings are stored
automatically and will be recalled when Quick Start is next selected.
TSURGE
Quick start
Press
START
and the test routines begin to work.
Press
CHANGE
and the actual parameters can be
changed.
All function keys except F2 (manual trigger) can
Stop
the test routine.
V
= 2000 V
+/-
=
+
tr
= 1s
n
=
endl
tri
= Auto
START CHANGE
F1
F2
F3
F4
F5
F6
F7
Page 3
(START)
TSURGE
Quick start
The user can select the parameter with the
related function key and can change the value
with the knob. The cursor allows the user to
define the value of the digit which should be
changed (fast or slow change).
V
= 2000 V
+/-
=
+
tr
= 1s
n
=
endl
tri
= Auto
Uset =
2000V
Counter
STOP
0000043
F1
F2
F3
F4
F5
F6
F7
Pressing the ESC button will bring the user back to the previous level from where the test can be restarted with
new levels. After restart the actual test time is displayed. All functions keys except F2 (MAN TRIGGER) can stop
the test routine. The latest setting will be displayed.
Any pressing of a function key will indicate the functions START, CHANGE or CONTINUE. F3 will continue the
same test routine. Also the test time will continue running. If the user selects at first START or CHANGE the test
will be stopped completely.
Page 3
(Change)
TSURGE
Quickstart
Note: At polarity setting ALT it is necessary to
double the number of impulses. Example n=2 =>
one impulse positive and one impulse negative.
Voltage
V
: 250V
-
6000V
Events
n
: 1
-
30000
Repetition
Rep : 2s
-
999s
U Imp A +/- cpl tr
250 12 Ohm 0 + L+N+PE 1 1/2
F1
F2
F3
F4
F5
F6
F7