6. Press
to accept the change or HOME Key
to cancel.
Figure 6-23: Purge Time
6.4
Reference gas - suggested concentration
ranges
Calibration gas ranges are recommended to be between 80 percent and 100 percent of
the full range of the analyzer. Please contact your Rosemount Customer Care
Representative if you require further assistance at (
Reference Manual
Gas calibration
00809-0100-4244
January 2020
Reference Manual
63