14. Carefully swing open the hinged PCB tray to gain access to the analysis cell. The
second analysis cell fitted in a Full rack system was fully accessible once the outer
cover was removed.
Figure 7-20: Accessing the Analysis Cell
15. Using an Allen key, remove and retain the six socket head cap screws and washers
(D) used to secure the cell cover (C) on the analysis cell (B). Care
MUST
be taken to
ensure the O-Ring (A) does not get damaged.
16. It may be necessary to use a small flat blade screwdriver to lever the cell cover off.
With the cover removed, the mirrors are now exposed.
Troubleshooting and diagnostics
Reference Manual
January 2020
00809-0100-4244
88