Figure 7-21: Opening the Analysis Cell
A. Analysis cell
B. O-ring
C. Cell cover
D. Fasteners
17. Clean the mirrors by applying a small quantity of reagent grade methanol to a
clean, dry, lens tissue and then gently wiping the mirrors with the lens tissue. Care
MUST
be taken not to apply pressure to the lens during this operation to prevent
scratching the reflective surface.
Scratches to the mirror surface
WILL
affect the ability for the analyzer to operate
and
WILL NOT
be covered by the warranty for the analyzer. (If reagent grade
methanol is unavailable, then reagent grade IPA may be used instead.)
Reference Manual
Troubleshooting and diagnostics
00809-0100-4244