Figure 1-4: Half Rack Dimensions
Figure 1-5: Full Rack Dimensions
Table 1-4: Environmental Characteristics
Environmental characteristic
Value
Units
Comment
Operating temperature range
5
℃
to 50
℃
41
℉
to 122
℉
℃
℉
Ambient temperature
Sample gas temperature range
15
℃
to 50
℃
59
℉
to 122
℉
℃
℉
N/A
Sample gas Pressure
0 - 1
atm (abs)
Typically 1 atm (abs)
Sample gas flow rate
1 - 5
L/min
N/A
Sample gas moisture content
2
%
Ensure the dew point of the
gas is at minimum 18
℉
(10
℃
) below the ambient
temperature to avoid
condensate in the gas lines
Sample gas particulate density
5
mg/m
3
Maximum
Sample gas particulate size
10
µm
Maximum
Maximum operating altitude
2000
m
N/A
Maximum relative humidity
80 @ 31
℃
%
Decreasing linearly to 50 %
relative humidity at 104
℉
(40
℃
)
Reference Manual
Plan
00809-0100-4244
January 2020
Reference Manual
19
Summary of Contents for Rosemount CT4400
Page 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Page 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Page 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Page 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Page 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Page 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Page 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Page 105: ......