7.7
Cleaning the analysis cell mirrors
Cleaning the analysis cell mirrors will only be required if the sample gas is not conditioned
in accordance with gas requirements defined in the General characteristics
and
the Environmental characteristics
.
For example, it is possible that the mirrors inside the cell may become contaminated if
contaminants are carried into the cell in water droplets in the gas being sampled if it is not
conditioned correctly the mirrors would then require to be cleaned to remove any
contamination from them.
The mirror surface is highly reflective and susceptible to damage if cleaned incorrectly or
excessively. Damage to the reflective coating on the mirrors
WILL
affect the analyzer
performance. Mirrors found to be damaged due to excessive cleaning or from corrosive
material residue will invalidate the warranty for the analyzer.
If the cell mirrors must be cleaned the analysis cell must only be opened in a
CLEAN/DRY/
DUST FREE
area to prevent further contamination.
Observe all safety precautions before starting this procedure.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (
Death or personal injury may result if this is not observed.
DANGER
FAILURE TO LOCK-OUT THE GAS HANDLING SYSTEM
Always lock out the gas handling system when shutting down the analyzer. Unauthorized
operation of the gas handling system when maintenance is being performed on the
analyzer or its associated pipes/hoses may result in highly flammable gas being released,
causing fire or explosion.
Death or personal injury may result if this is not observed.
Reference Manual
Troubleshooting and diagnostics
00809-0100-4244
January 2020
Reference Manual
85
Summary of Contents for Rosemount CT4400
Page 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Page 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Page 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Page 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Page 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Page 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Page 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Page 105: ......