11. Allow the concentration readings to stabilize.
Figure 5-5: Result
This screen gives a reading of the concentration of the selected gas that is present
as an impurity in the nitrogen calibration gas.
In the example the gas being measured is NH
3
, and the analyzer has detected a
concentration of 0.866 ppm.
If the reading is within tolerance, no further action is required.
12. Press
to end the zero verification process.
13. Press the appropriate button.
Option
Description
To return to the
Gas Service Screen
.
To select the next gas from the
Select
Gas screen
.
To return to the
Main Menu
.
To return to the
Gas Sensor Main
screen
).
Reference Manual
Verifying gas concentrations
00809-0100-4244
January 2020
Reference Manual
43