Option
Description
To select the next gas from the
Select
Gas screen
.
To return to the
Main Menu
.
To return to the
Gas Sensor Main
screen
(
Figure 6-9: Next Gas Service
6.3.2
Span calibration
The span gas concentrations measured by the analyzer when reference gas is flowing can
be verified and, if necessary, calibrated against the known reference gas by using the
following procedure.
Prerequisites
Run the analyzer at a stable temperature for at least thirty minutes prior to commencing
this procedure.
Procedure
1. Use a certified reference gas cylinder as the source of the span calibration gas.
2. Ensure that a suitably rated pressure regulator is connected to the reference gas
bottle.
3. Connect a hose from the reference gas bottle, through a T-piece, to the sample
supply port on the rear panel of the analyzer.
4. Connect an excess flow line to the unused port on the T-piece and route the excess
flow line to a suitable extractor.
5. On the display controller of the analyzer select the
Main Menu
screen.
Reference Manual
Gas calibration
00809-0100-4244
January 2020
Reference Manual
55
Summary of Contents for Rosemount CT4400
Page 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Page 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Page 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Page 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Page 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Page 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Page 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Page 105: ......