Figure 2-4: Gas Inlet and Outlet Connections
A. Rear panel
B. Sample gas outlet
C. Sample gas inlet
2.3
Connecting the electrical/electronic inputs and
outputs
This section discusses making the electrical connections to the analyzer.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (
Death or personal injury may result if this is not observed.
Install
Reference Manual
January 2020
00809-0100-4244
26
Rosemount CT4400
Summary of Contents for Rosemount CT4400
Page 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Page 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Page 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Page 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Page 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Page 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Page 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Page 105: ......