Instruction Manual
748441-D
April 2002
Rosemount Analytical Inc. A Division of Emerson Process Management
Serial Interface Option 4-5
Model CAT100
a. Status
Strings
If string syntax from the host computer is
faulty, or the analyzer is unable to act on
a received instruction, the analyzer will
transmit a status string to the host com-
puter as follows:
Status String Description
$ID;000;S100;LPB<CR> Unrecognized
instruction
code
$ID;000;S101;LPB<CR> LP-byte
in
error
$ID;000;S102;LPB<CR> Start
character
missing
$ID;000;S103;LPB<CR>
Input buffer overflow
$ID;xxx;S104;LPB<CR>
Analyzer off-line status
$ID;xxx;S105;LPB<CR> Text
line
too
long
$ID;xxx;S106;LPB<CR> Undefined
instruction
$ID;xxx;S107;LPB<CR> Invalid
integer
value
$ID;xxx;S108;LPB<CR>
Numerical value outside defined range
$ID;xxx;S109;LPB<CR> Invalid
failure/status
code
$ID;xxx;S110;LPB<CR>
Instruction cannot be done here
$ID;xxx;S111;LPB<CR>
Failure in transmitted character
$ID;xxx;S112;LPB<CR> Zeroing
running
$ID;xxx;S113;LPB<CR> Spanning
running
$ID;xxx;S114;LPB<CR>
Invalid real number
$ID;xxx;S115;LPB<CR> Automatic
calibration
mode
off
$ID;xxx;S116;LPB<CR>
Parameter outside defined range
$ID;xxx;S117;LPB<CR>
Pre-Flushing period is running
Where:
xxx:
Instruction code
ID:
Device ID number in RS485 mode
LPB: Message-length
parity
byte
<CR>: Terminate
character