Instruction Manual
748441-D
April 2002
4-8 Serial Interface Option
Rosemount Analytical Inc. A Division of Emerson Process Management
Model CAT100
b.
Response String Syntax
Response strings follow with the same syntax as the appropriate (SP-) commands.
The response string for instruction:
$ID;030;LPB<CR> SP Status messages
Is as follows:
$ID;030;a;b;c;LPB<CR>
Where:
Value Meaning
a: OK-Status
0
Relay without power
1
Relay
active
b: Calibration
0
No calibration
1
Zeroing
channel
1
2
Zeroing
channel
2
3
Zeroing channel 1 + 2
4
Spanning channel 1
5
Spanning channel 2
6
Spanning channel 1 + 2
7
Spanning channel 1 first, then channel 2
8
Reserved
9
Reserved
10
Waiting for flushing time and t90 response time
c: Relay 3
0
Relay without power
1
Relay
active