Instruction Manual
748441-D
April 2002
Rosemount Analytical Inc. A Division of Emerson Process Management
Description and Specifications 1-11
Model CAT100
1-4 SPECIFICATIONS
a. General
Power ............................................ Universal Power Supply 90-264 VAC, 50-60 Hz,
±
10% 180 Watts
Maximum at Start Up. Up to 380 Watts with optional case heater.
Channels (Detectors)/Number....... NDIR, PMD (PO
2
), E0
2
, TC. Up to two in one analyzer.
Mounting........................................ 4” or 6” Pipe, Rack, or Wall Mount
Area Classification......................... See Compliances page P-6
Corrosion Protection Option .......... Instrument grade air is required. Consult factory for requirements
Ambient Range.............................. Temperature (Standard): +5 °C to +45 °C (+41
°
to 113
°
F)
Temperature (Optional): -30 °C to +50 °C (-34
°
to 122
°
F)
Relative Humidity: 5 % to 95 %
Inputs/Outputs ............................... Digital: RS 232/ RS 485 serial data
or
FOUNDATION Fieldbus
Analog Current Outputs: Up to 2 isolated 4-20 mA, common
ground, burden: R
B
< 500 ohms
Analog Digital Outputs: Up to 8, 5-30 VDC, max current 30 mA
Analog Current/Voltage Input: Up to 3 (Option for TC electronic
cross compensation) 0(2) - 10 V, 0(4) - 20 mA, 0(0.2) - 1 V
Relay Outputs: Up to 3 “non-voltage carrying contacts”, max 30 V,
1 A, 30 W
Instrument Weight ......................... 120 to 150 lbs. (55-70 kg)