Instruction Manual
748441-D
April 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-11
Model CAT100
The Reset function restores the ana-
lyzer settings to the parameters and
calibration factors set in the factory.
This is the equivalent of switching off
the power and disabling the battery re-
tention of the memory by removing the
battery jumper J7.
NOTE:
All parameters and calibration fac-
tors entered by the user will be lost
whenever a reset operation is per-
formed
The currently valid user password must
be entered before a reset will be exe-
cuted.
Use the INPUT-CONTROL pads to en-
ter the password code and then acti-
vate ENTER.
Whenever a reset operation is per-
formed, the analyzer operating program
will be restarted, just as it is when the
analyzer is first switched on.
NOTE:
Jumper J6 on circuit board BKS,
which activates the watchdog cir-
cuitry must be inserted if the reset
operation is to be correctly exe-
cuted. See Section 5-5a, page 5-29.
o. Program
Version
This function displays the installed
software version number. This number
is also displayed when the analyzer is
switched on.
Activate the ENTER pad to continue.
p. Serial
Number
This function displays the analyzer se-
rial number.
Activate the ENTER pad for the con-
tinuation of the serial number.
Activate the FUNCTION pad until the
analyzer is back in the analysis mode.
q. Pump
This function turns the internal pump on
or off. The state of the pump is retained
when the power is shut off.
Entry of 0: The pump is off.
Entry of 1: The pump is on.