Model CAT100
f.
FUNCTION Touch pad
Activating the FUNCTION pad ad-
dresses the individual analyzer Func-
tions in sequence. Addressing an
analyzer Function will not initiate any
analyzer action or operation until the
ENTER pad is activated at the comple-
tion of an Input Control entry. The ana-
lyzer will continue to perform analysis
throughout touch pad entry procedures.
The following are the analyzer Func-
tions and their sequences:
Zeroing channel 1
Zeroing channel 2
Spanning channel 1
Spanning channel 2
Interval Time for automatic Zeroing
1
Interval Time for automatic Spanning
1
Entry of concentration limits
Entry of system parameters
Entry of serial interface parameters
2
Figure 3-6. CAT 100 Touch pad Functions
1
Only with combination of digital outputs and external solenoid valves, and if Auto = 1.
2
Only with optional RS 232 C/ RS 485 Serial Interface.