5-32 Maintenance and Service
Rosemount Analytical Inc. A Division of Emerson Process Management
Model CAT100
Analog Preamplifier (IR measurement)
Test Point:
X 25 channel 1 (analyzer
without oxygen meas-
urement)
X27 channel 2
Test Device:
DVM
Signal:
At zero gas purge: 0 V
DC (
±
100 mV)
There should be a mini-
mum difference of 600
mV (measuring range
<1000 ppm: difference
500 mV) between zero
point voltage and
sensitivity voltage.
Failure:
No signal or incorrect
measuring values
Possible rea-
sons:
Detector sensor not con-
nected
Detector sensor faulty
(replace detector)
BKS PCB faulty (replace
BKS)