Instruction Manual
748441-D
April 2002
Rosemount Analytical Inc. A Division of Emerson Process Management
Maintenance and Service 5-31
Model CAT100
Light Barrier Signal
Test Point:
Plug 9, pin 2
Test Device:
Oscilloscope
Signal:
Square wave U=6 V
P-P
(
±
0.3 V) Frequency = 24 Hz
(
±
0.1 Hz)
Failure: No
signal
Possible reasons (IR measurement):
Chopper not connected
Chopper inoperative
(switch analyzer off and
then on again)
Broken cable to light bar-
rier or faulty light barrier
(replace chopper)
BKS PCB faulty (replace
BKS)
Possible reasons (oxygen measurement
without infrared channel):
Solder bridge LB 18 not
closed
BKS PCB faulty (replace
BKS)
Analog Preamplifier (paramagnetic
oxygen measurement)
Test Point:
X 25 channel 1
Test Device:
DVM
Signal:
At zero gas purge: 0 V
DC (
±
50 mV)
At ambient air (21% O
2
):
500 mV DC (
±
50 mV)
Failure:
No signal or incorrect
measuring values
Possible rea-
sons:
Oxygen sensor not con-
nected Oxygen sensor
faulty (replace sensor)
BKS PCB faulty (replace
BKS)