NOTICE
Always install a new corrugated washer whenever you remove the sensing cell from the probe.
Bad reference side cell electrode
A bad reference side cell electrode can cause an elevated O
2
reading. This fault is usually
indicated by a frequent
Calibration Recommended
alarm and increasing cell
impedance readings. A high cell impedance can be calibrated out, but if the impedance
continues to increase rapidly, the sensing cell must be replaced.
4.5.2
Probe passes calibration, O
2
still reads low
The diffusion element at the end of the probe is a passive filter. It plugs very slowly,
because there is no active flow being drawn across it. In applications that have a heavy
particulate loading (coal or wood fired boilers, cement and lime kilns, catalyst
regeneration, recovery boilers, etc.) this diffusion element will eventually plug.
NOTICE
It is important not to pressurize the sensing cell during calibrations by flowing excessive cal
gas against a plugged diffuser. Always use a two-stage regulator for setting calibration gas
pressure. Calibration flow rates should be set only when a new diffuser is installed. As the
diffuser plugs, do not adjust the flow rates upward.
4.5.3
How do I detect a plugged diffuser?
The O
2
cell's speed of response will degrade. The O
2
trend in the control room will become
smoother.
When calibrating, the calibration gas flow rate will be noted to be lower. Never readjust
this flow upwards to correct for a plugged diffuser. Adjust this flow only when a new
diffuser is installed.
Always note the time it takes for the cell to recover to the normal process value after the
cal gas is removed. As the diffuser plugs, this recovery time will get longer and longer. Use
a calibration record to record and track calibration response times.
The Rosemount 6888Xi Advanced Electronics is available with an enhanced software
feature to automatically characterize the rate of diffuser pluggage during a calibration
cycle.
4.5.4
Can I calibrate a badly plugged diffuser?
It may not be possible to immediately replace a plugged diffuser while the process is on-
line.
Troubleshooting
94
Rosemount 6888A
Summary of Contents for Rosemount 6888A
Page 4: ......
Page 8: ...Contents iv Rosemount 6888A ...
Page 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Page 63: ...Configuration startup and operation Reference Manual 55 ...
Page 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Page 65: ...Configuration startup and operation Reference Manual 57 ...
Page 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Page 67: ...Configuration startup and operation Reference Manual 59 ...
Page 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Page 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Page 71: ...Configuration startup and operation Reference Manual 63 ...
Page 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Page 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Page 75: ...Configuration startup and operation Reference Manual 67 ...
Page 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Page 77: ...Configuration startup and operation Reference Manual 69 ...
Page 79: ...Configuration startup and operation Reference Manual 71 ...
Page 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Page 81: ...Configuration startup and operation Reference Manual 73 ...
Page 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Page 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Page 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Page 104: ...Troubleshooting 96 Rosemount 6888A ...
Page 120: ...Maintenance and service 112 Rosemount 6888A ...
Page 128: ...Optional accessories 120 Rosemount 6888A ...
Page 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Page 135: ...Safety data Reference Manual 127 ...