Environmental specifications
(continued)
Table 1-4:
General purpose certifications
Xi ambient temperature limits
-20 to 50 °C (-4 to 122 °F)
Xi temperature limits as measured inside the electronics
housing
-20 to 70 °C (-4 to 158 °F)
(1) Reduced cell life can be expected if operated continously at temperatures above 705 °C (1300 °F). Optional bypass and jacket
accessories permit operation to 1050 °C (1922 °F).
Installation specifications - probe
Table 1-5:
Probe mounting flange
Vertical or horizontal - 2 in. 150# (121 m (4.75 in.) bolt cir-
cle)
Note
Flanges are flat-faced and for mounting only. Flanges are
not pressure-rated. A 2.5-in. diameter hole in the process
is required.
Spool piece PN 3D39761G02 is available to offset probe
electronics housing from hot duct work.
Many adapter flanges are available to mate to existing
flanges.
Probe lengths and approximate shipping weights
457 mm (18 in.) package
7.3 kg (16 lb)
0.91 m (3 ft) package
9.5 kg (21 lb)
1.83 m (6 ft) package
12.2 kg (27 lb)
2.74 m (9 ft) package
15.0 kg (33 lb)
3.66 m (12 ft) package
17.7 kg (39 lb)
Reference air (optional)
2 scfh (1 L/min), clean, dry, instrument-quality air (20.95%
O
2
), regulated to 5 psi (34 kPa)
Calibration
Semi-automatic or automatic
Calibration gases
0.4% O
2
and 8% O
2
, balance N
2
recommended. Instrument
air may be used as a high cal gas but is not recommended.
100% nitrogen cannot be used as the low cal gas.
Calibration gas line
91 m (300 ft) maximum length
Calibration gas flow
5 scfh (2.5 L/min)
Heater electrical power
120/140 Vac ± 10%, 50/60 Hz, 260/1020 VA max, 1/2-in. -
14 NPT conduit ports
Traditional architecture cable
61 m (200 ft) maximum length
Power consumption of probe heater
776 VA maximum during warm-up
Description and specifications
Reference Manual
15
Summary of Contents for Rosemount 6888A
Page 4: ......
Page 8: ...Contents iv Rosemount 6888A ...
Page 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Page 63: ...Configuration startup and operation Reference Manual 55 ...
Page 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Page 65: ...Configuration startup and operation Reference Manual 57 ...
Page 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Page 67: ...Configuration startup and operation Reference Manual 59 ...
Page 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Page 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Page 71: ...Configuration startup and operation Reference Manual 63 ...
Page 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Page 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Page 75: ...Configuration startup and operation Reference Manual 67 ...
Page 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Page 77: ...Configuration startup and operation Reference Manual 69 ...
Page 79: ...Configuration startup and operation Reference Manual 71 ...
Page 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Page 81: ...Configuration startup and operation Reference Manual 73 ...
Page 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Page 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Page 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Page 104: ...Troubleshooting 96 Rosemount 6888A ...
Page 120: ...Maintenance and service 112 Rosemount 6888A ...
Page 128: ...Optional accessories 120 Rosemount 6888A ...
Page 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Page 135: ...Safety data Reference Manual 127 ...