![Emerson Rosemount 6888A Reference Manual Download Page 43](http://html.mh-extra.com/html/emerson/rosemount-6888a/rosemount-6888a_reference-manual_104744043.webp)
2.3.4
Transmitter probe with integral autocal and HART
communications
This probe contains gas-switching solenoids so that the Rosemount 6888Xi Electronics can
control the introduction of calibration gases. Calibrations can be initiated via a calibration
recommended diagnostic, time since last calibration, manually via external dry contact,
HART communications, or from the Rosemount 6888Xi local operator interface keypad.
The integral autocal feature can only be implemented when the probe is used with a
Rosemount 6888Xi.
1.
Remove the two covers from the transmitter.
2.
Connect the line (L1 wire) to the L1 terminal, the neutral (L2) wire to the L2/N
terminal, and the ground wire to the ground lug.
The Rosemount 6888A accepts 120/240 Vac ±10%, 50/60 Hz. No setup is required.
3.
Connect the 4-20 mA signal wires form the Rosemount 6888Xi to the connections in
the side chamber of the transmitter.
Do not connect the signal wires to the terminals in the main chamber where the AC
input wires are connected. Use a shielded twisted wire pair. Do not allow bare shield
wires to contact the circuit boards. Insulate the shield wires prior to termination. The
24 Vdc loop power is sourced from the Rosemount 6888Xi.
4.
Terminate the shield at both the probe and the Rosemount 6888Xi Advanced
Electronics.
NOTICE
The 4-20 mA signal represents the O
2
value and also powers the probe-mounted
electronics. Superimposed on the 4-20 mA signal is HART information accessible
through a field communicator or AMS software.
5.
Reinstall both covers on transmitter.
6.
Follow the remaining electrical installation instructions for the Rosemount 6888Xi
included with your system configuration.
See
Table 2-4
for the Rosemount 6888Xi product matrix.
Install
Reference Manual
35
Summary of Contents for Rosemount 6888A
Page 4: ......
Page 8: ...Contents iv Rosemount 6888A ...
Page 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Page 63: ...Configuration startup and operation Reference Manual 55 ...
Page 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Page 65: ...Configuration startup and operation Reference Manual 57 ...
Page 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Page 67: ...Configuration startup and operation Reference Manual 59 ...
Page 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Page 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Page 71: ...Configuration startup and operation Reference Manual 63 ...
Page 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Page 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Page 75: ...Configuration startup and operation Reference Manual 67 ...
Page 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Page 77: ...Configuration startup and operation Reference Manual 69 ...
Page 79: ...Configuration startup and operation Reference Manual 71 ...
Page 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Page 81: ...Configuration startup and operation Reference Manual 73 ...
Page 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Page 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Page 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Page 104: ...Troubleshooting 96 Rosemount 6888A ...
Page 120: ...Maintenance and service 112 Rosemount 6888A ...
Page 128: ...Optional accessories 120 Rosemount 6888A ...
Page 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Page 135: ...Safety data Reference Manual 127 ...