System Parameters
(continued)
Table 3-1:
Parameter name
Unit
Description
Alarm Level
--
O
2
alarm level
0
= 3.5 mA
1
= 21.1 mA
Cal Gas 1
%
Test gas 1 value. This is the actual value of the gas being
applied during the Test Gas 1 phase of a calibration.
Cal Gas 2
%
Test gas 2 value. This is the actual value of the gas being
applied during the Test Gas 2 phase of a calibration.
Gas Time
sec
Test gas application time. This is the length of time test
gases are applied to the O
2
probe during low or high test
gas phase of a calibration.
Purge Time
sec
Test gas purge time. This is the length of time before the
output will be returned to the process reading after a cali-
bration.
Slope
mV/Dec
Current calibration slope. This is the slope value that was
calculated as a result of the last successful calibration.
Constant
mV
Current calibration constant. This is the constant value
that was calculated as a result of the last successful calibra-
tion.
Impedance
ohm
Cell impedance. This is the sensor resistance that was cal-
culated as a result of the last successful calibration.
Time
Days ago
Time stamp of the last successful calibration.
Cal Logs | Slope
mV/Dec
Previous calibration slope. There are ten calibration results.
1
is the most recent, and
10
is the least recent calibration
slope.
Cal Logs | Constant
mV
Previous calibration constant. There are ten calibration re-
sults.
1
is the most recent, and
10
is the least recent cali-
bration constant.
Cal Logs | Impedance
ohm
Previous cell impedance. This is the sensor resistance that
was calculated as a result of a previous successful calibra-
tion. There are ten calibration results.
Index 1
is the
most recent, and
Index 10
is the least recent sensor re-
sistance measured.
Cal Logs | Time
Days ago
Time stamp of the previous successful calibration. There
are ten calibration time stamps.
Index 1
is the most re-
cent, and
Index 10
is the least recent time stamp.
Configuration, startup, and operation
78
Rosemount 6888A
Summary of Contents for Rosemount 6888A
Page 4: ......
Page 8: ...Contents iv Rosemount 6888A ...
Page 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Page 63: ...Configuration startup and operation Reference Manual 55 ...
Page 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Page 65: ...Configuration startup and operation Reference Manual 57 ...
Page 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Page 67: ...Configuration startup and operation Reference Manual 59 ...
Page 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Page 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Page 71: ...Configuration startup and operation Reference Manual 63 ...
Page 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Page 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Page 75: ...Configuration startup and operation Reference Manual 67 ...
Page 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Page 77: ...Configuration startup and operation Reference Manual 69 ...
Page 79: ...Configuration startup and operation Reference Manual 71 ...
Page 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Page 81: ...Configuration startup and operation Reference Manual 73 ...
Page 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Page 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Page 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Page 104: ...Troubleshooting 96 Rosemount 6888A ...
Page 120: ...Maintenance and service 112 Rosemount 6888A ...
Page 128: ...Optional accessories 120 Rosemount 6888A ...
Page 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Page 135: ...Safety data Reference Manual 127 ...