![Emerson Rosemount 6888A Reference Manual Download Page 101](http://html.mh-extra.com/html/emerson/rosemount-6888a/rosemount-6888a_reference-manual_104744101.webp)
4.5.1
Probe passes calibration, O
2
still reads high
External reference air leak
There may be a leak that is permitting ambient air to mix with the process gases. Since
many combustion processes are slightly negative in pressure, ambient air can be drawn
into the cell area, biasing the O
2
reading upward.
1.
Make sure that the calibration gas is capped tightly between calibrations.
2.
If autocal is used, make sure the check valve is sealing properly.
Probe Leakage Paths
Figure 4-2:
A. Cal gas 1/4 in. tube
B. Ref gas 1/4 in. tube
C. Vent
3.
If an abrasive shield is uses, a small leak at the probe flange can migrate down the
annular space between the probe OD and shield ID to the sensing cell, causing a
false high O
2
reading.
Internal reference air leak
See
Figure 4-2
. There may be a leak inside the O
2
probe itself, permitting the reference air
(20.95% O
2
) to mix with the process gases at the cell. To confirm this leak condition,
pressurize the inside (reference side) of the probe by plugging the reference air exhaust
port with your finger for one minute. (The conduit ports where the signal and power wires
pass may also need to be sealed.) The O
2
reading should decrease slightly. If the O
2
reading increases during this test, there is a leak inside the probe.
1.
Acid condensation inside the probe can degrade the hose that carries the cal gas to
the cell. Inspect this hose. Dislodging or improper installation of the cal gas or
reference air hose can cause a leakage path.
2.
The sensing cell is fastened to the end of the probe tube and uses a corrugated
washer to separate the process gases from the ambient reference air. The
corrugated washer may be damaged by corrosion. Discard used washer.
Troubleshooting
Reference Manual
93
Summary of Contents for Rosemount 6888A
Page 4: ......
Page 8: ...Contents iv Rosemount 6888A ...
Page 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Page 63: ...Configuration startup and operation Reference Manual 55 ...
Page 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Page 65: ...Configuration startup and operation Reference Manual 57 ...
Page 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Page 67: ...Configuration startup and operation Reference Manual 59 ...
Page 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Page 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Page 71: ...Configuration startup and operation Reference Manual 63 ...
Page 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Page 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Page 75: ...Configuration startup and operation Reference Manual 67 ...
Page 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Page 77: ...Configuration startup and operation Reference Manual 69 ...
Page 79: ...Configuration startup and operation Reference Manual 71 ...
Page 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Page 81: ...Configuration startup and operation Reference Manual 73 ...
Page 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Page 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Page 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Page 104: ...Troubleshooting 96 Rosemount 6888A ...
Page 120: ...Maintenance and service 112 Rosemount 6888A ...
Page 128: ...Optional accessories 120 Rosemount 6888A ...
Page 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Page 135: ...Safety data Reference Manual 127 ...