Reset the calibration as follows:
WARNING!
Failure to remove the Rosemount 6888A from automatic control loops prior to performing this
procedure may result in a dangerous operating condition.
Using the Field Communicator: Configure > Calibration > Calibration Constants > Reset
Calibration Constants.
Follow the instructions on the screen to complete the reset function.
3.11
D/A trim
The D/A trim procedure is used the calibrate the 4-20 mA output signal to a precision mA
measurement device. Only the signal to the DCS needs to be trimmed.
1.
Use the 375/475 Field Communicator to access the main HART menu.
2.
From the main menu, select
SERVICE TOOLS
.
3.
From the SERVICE TOOLS menu, select
MAINTENANCE
.
4.
From the MAINTENANCE menu, select
ANALOG OUTPUT
.
5.
From the ANALOG OUTPUT menu, select
TRIM mA OUTPUT
.
In the first screen, a
Loop should be removed from automatic control
warning occurs.
WARNING!
Failure to remove the Rosemount 6888A from automatic control loops prior to
performing this procedure may result in a dangerous operating condition.
6.
Remove the Rosemount 6888A from any automatic control loops to avoid a
potentially dangerous operating condition. Press OK when ready.
7.
Follow the prompts on the 375/475 to perform the trim.
8.
Use the stylus to enter the new measured analog output value; then press ENTER to
accept.
9.
When trim is complete, press OK to return to the ANALOG OUTPUT menu.
Configuration, startup, and operation
84
Rosemount 6888A
Summary of Contents for Rosemount 6888A
Page 4: ......
Page 8: ...Contents iv Rosemount 6888A ...
Page 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Page 63: ...Configuration startup and operation Reference Manual 55 ...
Page 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Page 65: ...Configuration startup and operation Reference Manual 57 ...
Page 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Page 67: ...Configuration startup and operation Reference Manual 59 ...
Page 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Page 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Page 71: ...Configuration startup and operation Reference Manual 63 ...
Page 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Page 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Page 75: ...Configuration startup and operation Reference Manual 67 ...
Page 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Page 77: ...Configuration startup and operation Reference Manual 69 ...
Page 79: ...Configuration startup and operation Reference Manual 71 ...
Page 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Page 81: ...Configuration startup and operation Reference Manual 73 ...
Page 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Page 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Page 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Page 104: ...Troubleshooting 96 Rosemount 6888A ...
Page 120: ...Maintenance and service 112 Rosemount 6888A ...
Page 128: ...Optional accessories 120 Rosemount 6888A ...
Page 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Page 135: ...Safety data Reference Manual 127 ...