If there is an error condition at startup, an alarm message is displayed on the Rosemount
6888Xi.
3.4
Rosemount 6888Xi Quick Start Wizard
When the Rosemount 6888Xi is first powered, a short wizard program guides you through
the basic setup. Once configured, the Rosemount 6888Xi retains the setup, and the wizard
will not repeat.
1.
Apply power to the Rosemount 6888Xi.
Once boot-up is complete, the Quick Start Wizard screen appears. With a dual
channel Rosemount 6888Xi, the wizard runs for both channels in succession.
2.
Press Enter to continue.
3.
At the Sensor Type screen, use the Up and Down keys to select O
2
.
Do not select
CO
as this option is reserved for future use.
4.
Press Enter to continue.
5.
At the Device Type screen use the Up and Down keys to select HART
®
or FF
(FOUNDATION
™
Fieldbus), whichever applies.
6.
At the Auto Cal Device screen, use the Up and Down keys to select the calibration
method to be used. The methods are defined as follows:
•
None
- Manual calibration with the standard probe housing configuration
•
SPS
- Automatic calibration with the standard probe housing configuration using
the Rosemount SPS 4001B
•
IMPS
- Automatic calibration with the standard probe housing configuration
using the Rosemount IMPS
•
Integral
- Automatic calibration with the integral autocal probe housing
configuration
7.
Press Enter to continue.
NOTICE
If
SPS
,
IMPS
, or
Integral
is selected, you must still configure automatic calibration as
On
.
Other parameters, such as test gas values and gas times, should be verified as well. Refer
to the
Rosemount 6888Xi Reference Manual
for calibration setup details.
8.
When prompted by
Setup Correct?
, use the Up and Down keys to select
Yes
.
If you select
No
, the wizard restarts.
9.
Press Enter to continue.
The Rosemount 6888Xi displays several screens while saving the configuration, resets
itself, and then returns to the main screen.
Configuration, startup, and operation
Reference Manual
49
Summary of Contents for Rosemount 6888A
Page 4: ......
Page 8: ...Contents iv Rosemount 6888A ...
Page 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Page 63: ...Configuration startup and operation Reference Manual 55 ...
Page 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Page 65: ...Configuration startup and operation Reference Manual 57 ...
Page 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Page 67: ...Configuration startup and operation Reference Manual 59 ...
Page 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Page 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Page 71: ...Configuration startup and operation Reference Manual 63 ...
Page 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Page 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Page 75: ...Configuration startup and operation Reference Manual 67 ...
Page 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Page 77: ...Configuration startup and operation Reference Manual 69 ...
Page 79: ...Configuration startup and operation Reference Manual 71 ...
Page 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Page 81: ...Configuration startup and operation Reference Manual 73 ...
Page 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Page 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Page 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Page 104: ...Troubleshooting 96 Rosemount 6888A ...
Page 120: ...Maintenance and service 112 Rosemount 6888A ...
Page 128: ...Optional accessories 120 Rosemount 6888A ...
Page 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Page 135: ...Safety data Reference Manual 127 ...