NOTICE
The optional Rosemount SPS 4001B or Rosemount IMPS 4000 Sequencer can only be used
when the Rosemount 6888Xi advanced electronics option is selected. The Rosemount 6888Xi
must be properly configured for autocalibration. See
Chapter 3
.
See the
SPS 4001B Single Probe Autocalibration Sequencer Instruction Manual
or the
IMPS 4000 Intelligent Multiprobe Test Gas Sequencer Instruction Manual
for wiring and
pneumatic connections.
2.4.2
Calibration gas
Two calibration gas concentrations are used with this transmitter, low gas - 0.4% O
2
,
balance N
2
, and high gas - 8% O
2
, balance N
2
.
CAUTION!
CALIBRATION READING ERRORS
Do not use 100% nitrogen as a low gas (zero gas). Emerson suggests that gas for the low (zero)
be between 0.4% and 2.0% O
2
. Do not use gases with hydrocarbon concentrations of more than
40 parts per million. Failure to use proper gases will result in erroneous readings.
CAUTION!
EQUIPMENT DAMAGE
If the ducts will be washed down during outage, MAKE SURE to power down the Rosemount
6888A units and remove them from the wash areas.
NOTICE
Upon completing installation, make sure that the transmitter is turned on and operating prior
to firing up the combustion process. Damage can result from having a cold unit exposed to the
process gases. During outages, if possible, leave all units running to prevent condensation and
premature aging from thermal cycling.
An optional manual calibration panel is shown in
Figure 2-17
. See
Figure 2-18
for the
Rosemount 6888A probe calibration gas connection ports.
Install
Reference Manual
45
Summary of Contents for Rosemount 6888A
Page 4: ......
Page 8: ...Contents iv Rosemount 6888A ...
Page 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Page 63: ...Configuration startup and operation Reference Manual 55 ...
Page 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Page 65: ...Configuration startup and operation Reference Manual 57 ...
Page 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Page 67: ...Configuration startup and operation Reference Manual 59 ...
Page 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Page 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Page 71: ...Configuration startup and operation Reference Manual 63 ...
Page 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Page 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Page 75: ...Configuration startup and operation Reference Manual 67 ...
Page 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Page 77: ...Configuration startup and operation Reference Manual 69 ...
Page 79: ...Configuration startup and operation Reference Manual 71 ...
Page 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Page 81: ...Configuration startup and operation Reference Manual 73 ...
Page 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Page 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Page 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Page 104: ...Troubleshooting 96 Rosemount 6888A ...
Page 120: ...Maintenance and service 112 Rosemount 6888A ...
Page 128: ...Optional accessories 120 Rosemount 6888A ...
Page 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Page 135: ...Safety data Reference Manual 127 ...