![Emerson Rosemount 6888A Reference Manual Download Page 90](http://html.mh-extra.com/html/emerson/rosemount-6888a/rosemount-6888a_reference-manual_104744090.webp)
3.10.1
Calibration procedure
This section covers manual calibration. For automatic calibration details, see the
instruction manual for the Rosemount SPS 4001B Single Probe Autocalibration Sequencer
or the Rosemount IMPS 4000 Intelligent Multiprobe Test Gas Sequencer.
1.
Use the 375/475 Field Communicator to access the main HART menu.
2.
From the main menu, select
CONFIGURE
.
3.
From the CONFIGURE menu, select
CALIBRATION
.
4.
From the CALIBRATION menu, again select
CALIBRATION
.
5.
From the CALIBRATION menu, select O
2
CALIBRATION.
WARNING!
Failure to remove the Rosemount 6888A from automatic control loops prior to
performing this procedure may result in a dangerous operating condition.
Calibration Method - Simplified
Figure 3-6:
6.
In the first screen, a
Loop should be removed from automatic control
warning occurs. Remove the Rosemount 6888A from any automatic loop controls to
avoid a potentially dangerous operating condition. Press OK when ready.
7.
At the next screen, when the step shows
APPLY GAS 1
, press OK to continue.
8.
When Flow Gas 1 and Read Gas 1 are complete and the step shows
APPLY GAS 2
,
press
OK
to continue.
9.
When Flow Gas 2 and Read Gas 2 are complete and the step shows
STOP GAS
,
press OK to continue the calibration with purge.
10.
When the screen shows
Loop may be returned to automatic control
,
press OK to return to the CALIBRATION screen.
11.
At the CALIBRATION menu, select
RESULT
.
At the RESULT menu, the results of the calibration are displayed. In the event the
calibration cycle fails, the reason is displayed here as well. The calibration results are
described as follows:
Configuration, startup, and operation
82
Rosemount 6888A
Summary of Contents for Rosemount 6888A
Page 4: ......
Page 8: ...Contents iv Rosemount 6888A ...
Page 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Page 63: ...Configuration startup and operation Reference Manual 55 ...
Page 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Page 65: ...Configuration startup and operation Reference Manual 57 ...
Page 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Page 67: ...Configuration startup and operation Reference Manual 59 ...
Page 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Page 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Page 71: ...Configuration startup and operation Reference Manual 63 ...
Page 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Page 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Page 75: ...Configuration startup and operation Reference Manual 67 ...
Page 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Page 77: ...Configuration startup and operation Reference Manual 69 ...
Page 79: ...Configuration startup and operation Reference Manual 71 ...
Page 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Page 81: ...Configuration startup and operation Reference Manual 73 ...
Page 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Page 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Page 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Page 104: ...Troubleshooting 96 Rosemount 6888A ...
Page 120: ...Maintenance and service 112 Rosemount 6888A ...
Page 128: ...Optional accessories 120 Rosemount 6888A ...
Page 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Page 135: ...Safety data Reference Manual 127 ...