Configuration and Use Manual
57
Required Transmitter Configuration
Req
u
ir
ed
Con
fig
uration
Opt
ion
al Co
nfi
g
u
ratio
n
Using the
T
ransmitter
Flo
w
me
te
r Sta
rtup
Table 6-22
Discrete input polarity
Polarity
Input power supply
DI status
Description
Active high
Internal
ON
Voltage across terminals is high
OFF
Voltage across terminals is zero
External
ON
Voltage applied across terminals is 3–30 VDC
OFF
Voltage applied across terminals is <0.8 VDC
Active low
Internal
ON
Voltage across terminals is zero
OFF
Voltage across terminals is high
External
ON
Voltage applied across terminals is <0.8 VDC
OFF
Voltage applied across terminals is 3–30 VDC