Micro Motion
®
Model 2400S Transmitters with Analog Outputs
Measurement Performance
There are two versions of the meter verification application: the original version and Micro Motion
Smart Meter Verification. Table 10-1 lists requirements for each version. Table 10-2 provides a
comparison of the two versions.
Note: If you are running an older version of ProLink II or the Communicator device description, you
will not be able to access the additional features in Smart Meter Verification. If you are running an
updated version of ProLink II or the Communicator with the original version of meter verification, the
meter verification procedures will be slightly different from the procedures shown here.
Table 10-1
Version requirements for meter verification application
Requirement type
Meter verification application
Original version
Smart Meter Verification
Transmitter
v1.0
v4.0
ProLink II requirements
v2.5
v2.9
HART DD requirements
375 Field Communicator device rev 1,
DD rev 1
375 Field Communicator device rev 4,
DD rev 2
Table 10-2
Comparison of meter verification features and functions: original version vs. Smart Meter
Verification
Feature or function
Meter verification application
Original version
Smart Meter Verification
Process interruption
No need to halt flow
No need to halt flow
Measurement interruption
Three minutes. Outputs go to:
• Last Measured Value
• Configured Fault Value
User option:
• Continue Measurement. Measurement is
not interrupted. Test requires
approximately 90 seconds.
• Last Measured Value. Outputs fixed and
measurement interrupted for
approximately 140 seconds.
• Configured Fault Value Outputs fixed and
measurement interrupted for
approximately 140 seconds.
Result storage
Test results stored only for tests run with
ProLink II, and stored on the PC
Twenty most recent results stored on the
transmitter, independent of tool used to
perform the procedure. For tests run with
ProLink II, additional result data stored on
PC.
Result data on display
Pass/Fail/Abort for current test
For all results stored on transmitter:
• Pass/Fail/Abort
• Abort code (if relevant)
• Stiffness of the right and left pickoffs
Summary of Contents for Micro-Motion 2400S
Page 8: ...vi Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 20: ...12 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 30: ...22 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 42: ...34 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 66: ...58 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 76: ...68 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 138: ...130 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 190: ...182 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 197: ......