138
Micro Motion
®
Model 2400S Transmitters with Analog Outputs
Troubleshooting
Table 11-4
Status alarms and remedies
Alarm
code
Communicator
Cause
Suggested remedy
ProLink II
A001
EEprom Checksum
Error (Core
Processor)
An uncorrectable
checksum mismatch has
been detected.
• Cycle power to the flowmeter.
• The flowmeter might need service. Contact Micro
Motion. See Section 11.3.
(E)EPROM
Checksum Error
(CP)
A002
RAM Test Error
(Core Processor)
ROM checksum error or a
RAM location cannot be
written to.
• Cycle power to the flowmeter.
• The flowmeter might need service. Contact Micro
Motion. See Section 11.3.
RAM Error (CP)
A003
Sensor Not
Responding (No
Tube Interrupt)
Continuity failure of drive
circuit, LPO, or RPO, or
LPO-RPO mismatch when
driving.
• Check for slug flow. See Section 11.17.
• Check the test points. See Section 11.24.
• Check the sensor circuitry. See Section 11.25.
• Check sensor tubes for plugging.
• If the problem persists, contact Micro Motion. See
Sensor Failure
A004
Temperature
sensor out of range
Combination of A16 and
A17
• Check the sensor RTD circuitry. See Section 11.25.
• Verify that process temperature is within range of
sensor and transmitter.
• If the problem persists, contact Micro Motion. See
Temperature
Sensor Failure
A005
Input Over-Range
The measured flow has
exceeded the maximum
flow rate of the sensor
(
Δ
T > 200
μ
s)
• If other alarms are present (typically, A003, A006,
A008, A102, or A105), resolve those alarm
conditions first. If the A005 alarm persists, continue
with the suggestions here.
• Verify process and check for slug flow. See
• Check the test points. See Section 11.24.
• Check the sensor circuitry. See Section 11.25.
• Check the sensor tubes for erosion. See
• If the problem persists, contact Micro Motion. See
Input Overrange
A006
Transmitter Not
Characterized
Combination of A020 and
A021
• Check the characterization. Specifically, verify the
FCF and K1 values. See Section 6.2.
• If the problem persists, contact Micro Motion. See
Not Configured
A008
Density Outside
Limits
The measured density has
exceeded 0–10 g/cm
3
• If other alarms are present (typically, A003, A006,
A102, or A105), resolve those alarm conditions first.
If the A008 alarm persists, continue with the
suggestions here.
• Verify process. Check for air in the flow tubes, tubes
not filled, foreign material in tubes, or coating in tubes
(see Section 11.18).
• Check for slug flow. See Section 11.17.
• Check the sensor circuitry. See Section 11.25.
• Verify calibration factors in transmitter configuration.
• Check the test points. See Section 11.24.
• If the problem persists, contact Micro Motion. See
Density Overrange
Summary of Contents for Micro-Motion 2400S
Page 8: ...vi Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 20: ...12 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 30: ...22 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 42: ...34 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 66: ...58 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 76: ...68 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 138: ...130 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 190: ...182 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Page 197: ......