DVC6000f Digital Valve Controllers
December 2009
1-10
Table 1-5. Hazardous Area Classifications—ATEX
Certificate
Type
Certification Obtained
Entity Rating
Temperature
Code
Enclosure Rating
ATEX
DVC60x0F
DVC60x0FS
(x = 1,2,3)
II 1 G D
Gas
Ex ia IIC T4/T5/T6—Intrinsically Safe
Dust
Ex iaD 20 T100
C (Tamb
v
80
C)
Ex iaD 20 T85
C (Tamb
v
75
C)
FIELDBUS
T4(Tamb
v
80
C)
T5(Tamb
v
77
C)
T6(Tamb
v
62
C)
IP66
Ui = 24 VDC , Ii = 380 mA ,
Ci = 5 nF , Li = 0 mH , Pi = 1.4 W
FISCO
Ui = 17.5 VDC , Ii = 380 mA ,
Ci = 5 nF , Li = 0 mH , Pi = 5.32 W
II 2 G D
Gas
Ex d IIB+H2 T5/T6—Flameproof
Dust
Ex tD A21 IP66 T90
C (Tamb
v
85
C)
Ex tD A21 IP66 T80
C (Tamb
v
75
C)
−
−
−
T5(Tamb
v
85
C)
T6(Tamb
v
75
C)
IP66
II 3 G D
Gas
Ex nCL IIC T5/T6—Type n
Dust
Ex tD A22 IP66 T85
C (Tamb
v
80
C)
Ex tD A22 IP66 T80
C (Tamb
v
75
C)
−
−
−
T5(Tamb
v
80
C)
T6(Tamb
v
75
C)
IP66
DVC6005F
II 1 G D
Gas
Ex ia IIC T4/T5/T6—Intrinsically Safe
Dust
Ex iaD 20 T100
C (Tamb
v
80
C)
Ex iaD 20 T85
C (Tamb
v
75
C)
FIELDBUS
T4(Tamb
v
80
C)
T5(Tamb
v
77
C)
T6(Tamb
v
62
C)
IP66
Ui = 24 VDC
Ii = 380 mA
Ci = 5 nF
Li = 0 mH
Pi = 1.4 W
Uo = 24 VDC
Io = 17.5 mA
Co = 121 nF
Lo = 100 mH
Po = 105 mW
FISCO
Ui = 17.5 VDC
Ii = 380 mA
Ci = 5 nF
Li = 0 mH
Pi = 5.32 W
Uo = 24 VDC
Io = 17.5 mA
Co = 121 nF
Lo = 100 mH
Po = 105 mW
II 2 G D
Gas
Ex d IIB T5/T6—Flameproof
Dust
Ex tD A21 IP66 T85
C (Tamb
v
80
C)
Ex tD A21 IP66 T75
C (Tamb
v
70
C)
−
−
−
T5(Tamb
v
80
C)
T6(Tamb
v
70
C)
IP66
II 3 G D
Gas
Ex nL IIC T5/T6—Type n
Dust
Ex tD A22 IP66 T85
C (Tamb
v
80
C)
Ex tD A22 IP66 T80
C (Tamb
v
75
C)
−
−
−
T5(Tamb
v
80
C)
T6(Tamb
v
75
C)
IP66
DVC60x5
(x = 1,2,3)
II 1 G D
Gas
Ex ia IIC T4/T5/T6—Intrinsically Safe
Dust
Ex iaD 20 T135
C (Tamb
v
125
C)
Ex iaD 20 T100
C (Tamb
v
95
C)
Ex iaD 20 T85
C (Tamb
v
80
C)
Ui = 30 VDC
Ii = 17.5 mA
Ci = 0
μ
F
Li = 0 mH
Pi = 105 mW
T4(Tamb
v
125
C)
T5(Tamb
v
95
C)
T6(Tamb
v
80
C)
IP66
II 2 G D
Gas
Ex d IIC T4/T5/T6—Flameproof
Dust
Ex tD A21 IP66 T130
C (Tamb
v
125
C)
Ex tD A21 IP66 T100
C (Tamb
v
95
C)
Ex tD A21 IP66 T85
C (Tamb
v
80
C)
−
−
−
T4(Tamb
v
125
C)
T5(Tamb
v
95
C)
T6(Tamb
v
80
C)
IP66
II 3 G D
Gas
Ex nA IIC T4/T5/T6—Type n
Dust
Ex tD A22 IP66 T130
C (Tamb
v
125
C)
Ex tD A22 IP66 T100
C (Tamb
v
95
C)
Ex tD A22 IP66 T85
C (Tamb
v
80
C)
−
−
−
T4(Tamb
v
125
C)
T5(Tamb
v
95
C)
T6(Tamb
v
80
C)
IP66
1
Summary of Contents for Fisher FIELDVUE DVC6000f
Page 54: ...DVC6000f Digital Valve Controllers December 2009 2 30 2 ...
Page 62: ...DVC6000f Digital Valve Controllers December 2009 4 2 4 ...
Page 127: ...Transducer Block December 2009 4 67 4 ...
Page 141: ...AO Function Block December 2009 4 81 4 ...
Page 207: ...MAI Function Block December 2009 4 147 4 ...
Page 219: ...DO Function Block December 2009 4 159 4 ...
Page 231: ...DI Function Block December 2009 4 171 4 ...
Page 302: ...DVC6000f Digital Valve Controllers December 2009 8 16 8 ...
Page 306: ...DVC6000f Digital Valve Controllers December 2009 A 4 A ...
Page 350: ...DVC6000f Digital Valve Controllers December 2009 E 6 E ...
Page 368: ...DVC6000f Digital Valve Controllers December 2009 F 18 F ...
Page 372: ...DVC6000f Digital Valve Controllers December 2009 Glossary 4 Notes G Glossary ...
Page 382: ...DVC6000f Digital Valve Controllers December 2009 Index 10 F Index ...