3410 Series Gas Ultrasonic Meters Operations Manual
Section 7: Tools commands
3-9000-777 Rev D
June 2016
Error recovery for program download failure
137
unsuccessful, the meter may have defaulted to 192.168.135.100 with a subnet
of 255.255.255.0. Make sure your PC has a compatible address and attempt a
connection using this IP address.
2.
Once your cabling and Meter Directory record are setup, attempt the connection to the
meter. You will receive a message
Error 10001 opening database connection to
, click
OK
.
3.
At the prompt to
Attempt FTP-only connection
, Click
Yes
. If successful, the Daniel
MeterLink caption reads
…Connected to <meter name>
.
4.
Go to
Tools|Program Download
to attempt the firmware upgrade again. If the firmware
upgrade is successful, the meter should start working as normally as the meter’s
configuration is not normally lost.
5.
If the configuration is lost, a configuration file was saved by the Program Download
utility on the initial firmware upgrade attempt.
Use
Edit/Compare Configuration
to write this saved configuration back to the meter. The
saved configuration files are typically stored in
C:\Ultrasonic Data\<MeterName>
.
6.
If the firmware upgrade fails again, contact Daniel Technical Support for assistance.
Contact information can be found in Daniel MeterLink under the
Help
pull down menu
by selecting Technical Support. Or, in the Customer service in the preface of this
manual.