![ELTRA CS-800 Manual Download Page 93](http://html.mh-extra.com/html/eltra/cs-800/cs-800_manual_2397410093.webp)
93
-
Shift the switch with holder downwards until the magnetic switch is off (A
valve clicks and the piston slightly moves downwards).
-
Make a second marking on the cylinder (6) at the upper end of the switch
holder (4).
-
Adjust the switch holder (4) to have the upper end in the middle of the two
markings.
-
Open the furnace and then close it again (F2) to verify the correctness of the
adjustment.
-
Adjust the lower cushioning (7), so that the furnace opens without the piston
banging at the bottom.
-
Click on “Communication test”-“Digital outputs”, on the 9th button counting
from the left, to energize the coil cooling valve (11/V8), or start analysis
cycle.
-
Turn the restrictor (12) counter clockwise until the pressure on the gauge
(13) is reduced by about 0.2 to 0.3 bar (3 to 4 psi) when the coil cooling
valve (11/V8) is open.
-
Fix the locknut of the restrictor (12).
Remark:
In case of using valuable gas for operating the pneumatics, the restrictor (12)
should be adjusted for lower flow by turning it clockwise. It can even be completely
closed by turning it clockwise up to the end of its range.
The flow adjusted with this restrictor is for cooling the combustion coil to ensure
very long service life. According to our experience, the combustion coils of our
analyzers don’t need replacement, even after years of operation.
Pos : 7. 19 /0020 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Press ure regul ati on ELTR A @ 6\ mod_1369298860002_9. doc x @ 41930 @ 32233 @ 1
4.4
46B
Pressure regulation
Pos : 7. 20 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-800_Ser vic e/ 0020 Eins tell ungen, Tests und F unktions beschr eibung/0255 M odul Pres sur e R egulation @ 6\ mod_1377688630939_9.doc x @ 46975 @ 2424 @ 1
Fig. 48: PR1
PR1 is the inlet oxygen regulator, and is situated inside the analyzer, on the rear
panel.
The outlet pressure needs to be set to 1.5bar (22.5psi).
•
Pull and rotate the knob of the regulator, until the pressure gauge (6) shows
the correct value.
–
Ensure that the flow shown on the flow meter (5) shows between 180 and
200l/h; if not, adjust according to the related chapter.
Summary of Contents for CS-800
Page 3: ...3 ...
Page 10: ...10 Fig 2 Draft 12057 ...
Page 11: ...11 Fig 3 Draft 11660 ...
Page 12: ...12 Fig 4 Draft 12200 ...
Page 14: ...14 Fig 6 Draft 12058 Signal and control ...
Page 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Page 18: ...18 Fig 9 Oxygen flow schematically ...
Page 19: ...19 Fig 10 Draft 12059 signal and control board ...
Page 21: ...21 Fig 12 Draft VG 96 connector ...
Page 22: ...22 Fig 13 Draft 1 14 4 1 ...
Page 23: ...23 Fig 14 Draft IR 45 power supply ...
Page 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Page 80: ...80 Abb 46 IRC 1 3 ...
Page 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Page 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Page 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Page 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Page 110: ...Service 110 ...
Page 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Page 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Page 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Page 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Page 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Page 134: ...Miscellaneous 134 Fig 47 12055 I ...
Page 135: ...Miscellaneous 135 Fig 48 12055 II ...
Page 136: ...Miscellaneous 136 Fig 49 12072 ...
Page 137: ...Miscellaneous 137 Fig 50 12064 ...
Page 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Page 139: ...Miscellaneous 139 Fig 52 12015 I ...
Page 140: ...Miscellaneous 140 Fig 53 12015 II ...
Page 141: ...Miscellaneous 141 Fig 54 12016 ...
Page 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Page 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Page 145: ...Miscellaneous 145 ...
Page 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Page 155: ......
Page 156: ......
Page 157: ......