28
The IR cells have normally base lines adjusted to about 8V and the peak comes
down during analysis, respectively during IR absorption by the measuring gas.
In case of golden paths, although we intensively clean them with solvents and
heated up to over 500°C for one day or more, in the first couple of months after
being in operation, the base line may goes up, which means that the paths become
more reflective during this period. Therefore, the base lines of golden paths should
be initially set rather at 5 or 6 volts instead of 8 volts. In the first couple of months
the base lines go up, but the growing slows down and growing contamination
finally causes the base line to start coming down.
The initial base line adjustment at 5 to 6 volts, prevents the maximum base line
value from reaching 9.5V, where the error message “Check base line” appears.
Note that this message is just a warning. The analyzer is not prevented from
carrying out further analyses. As long as the base line remains below 10V, the
results are correct.
If a base line is below -9.5V an adjustment should be done as advised above. If
this is not possible, then most probably there is a fault or a very strong
contamination. In case of contamination, all cells available should have very low
voltages, even negative ones.
If all available cells are below -9.5V there is most probably a fault like no rotating
chopper motor or a faulty infrared source.
Pos : 5. 4 / 0020 Ü bersc hrift en/ELTR A 1. 1 Ü bersc hrift en/ 11 Fluc tuati ng gas flow ELTRA @ 6\ mod_1368452950593_9. doc x @ 40051 @ 22 @ 1
3.2
15B
Fluctuating gas flow
Pos : 5. 5 / 0010 ELTR A/ 0015 s er vice_i nstr uctions /CS- 800_Ser vice/0015 Fehler/ 0110 Bild Fluktui erender Gas fluss (Front ansic ht - NR 4, 5, 6, 7) ELTR A @ 6\ mod_1377688597933_9. doc x @ 46495 @ @ 1
Fig. 4: Fluctuating gas flow
Pos : 5. 6 / 0010 ELTR A/ 0015 s er vice_i nstr uctions /CS- 800_Ser vice/0015 Fehler/ 0115 M odul Fl ukt uier ender G asfl uss ELTRA @ 6\ mod_1377688598185_9. doc x @ 46507 @ 2223333333333333333 @ 1
The flow rate is electronically controlled; therefore there should be no significant
variation from the set value.
7
6
5
Summary of Contents for CS-800
Page 3: ...3 ...
Page 10: ...10 Fig 2 Draft 12057 ...
Page 11: ...11 Fig 3 Draft 11660 ...
Page 12: ...12 Fig 4 Draft 12200 ...
Page 14: ...14 Fig 6 Draft 12058 Signal and control ...
Page 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Page 18: ...18 Fig 9 Oxygen flow schematically ...
Page 19: ...19 Fig 10 Draft 12059 signal and control board ...
Page 21: ...21 Fig 12 Draft VG 96 connector ...
Page 22: ...22 Fig 13 Draft 1 14 4 1 ...
Page 23: ...23 Fig 14 Draft IR 45 power supply ...
Page 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Page 80: ...80 Abb 46 IRC 1 3 ...
Page 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Page 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Page 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Page 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Page 110: ...Service 110 ...
Page 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Page 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Page 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Page 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Page 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Page 134: ...Miscellaneous 134 Fig 47 12055 I ...
Page 135: ...Miscellaneous 135 Fig 48 12055 II ...
Page 136: ...Miscellaneous 136 Fig 49 12072 ...
Page 137: ...Miscellaneous 137 Fig 50 12064 ...
Page 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Page 139: ...Miscellaneous 139 Fig 52 12015 I ...
Page 140: ...Miscellaneous 140 Fig 53 12015 II ...
Page 141: ...Miscellaneous 141 Fig 54 12016 ...
Page 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Page 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Page 145: ...Miscellaneous 145 ...
Page 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Page 155: ......
Page 156: ......
Page 157: ......