Fig. 24: Pneumatics (photo)
Little or no compressed air pressure, the furnace cannot close.
This pressure is shown on gauge (13), on the front-side of the analyzer.
It must be between 4 to 6 bar (60 to 90psi).
The screw (2) must be turned counter-clockwise, as much as possible.
Adjust regulator (8), so that the furnace floor can rise completely, without shaking;
while a person still can stop the moving furnace floor with only one hand.
This is a safety precaution.
Repeat above instruction, until the adjustment is completed. Now, the upward
speed is fast enough, yet safe for the operator. The pressure regulator (8) is found
on the inside, on the right-side section of the analyzer, installed on the rear panel.
One of the pneumatic valves (V7) or (V10) (or both) is malfunctioning.
The magnetic coil of the pneumatic valve (V7) or (V10) is not powered. The
furnace floor goes up, when valve (V10) receives 24 VDC.
The 24 V DC comes from the microcontroller board; if not, replace the electronic
drivers.
Pos : 5. 19 /0020 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Anal ys edauer zu lang - The anal ysis t akes t oo long ELTR A @ 6\ mod_1368613082100_9.doc x @ 40314 @ 22 @ 1
3.8
21B
Analysis takes too long
Pos : 5. 20 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-800_Ser vic e/ 0015 F ehl er/ 0150 Modul Anal ys edauer z u l ang ELTRA @ 6\ mod_1377688604344_9. doc x @ 46591 @ 2224 @ 1
12
11
9
10
8
15
Summary of Contents for CS-800
Page 3: ...3 ...
Page 10: ...10 Fig 2 Draft 12057 ...
Page 11: ...11 Fig 3 Draft 11660 ...
Page 12: ...12 Fig 4 Draft 12200 ...
Page 14: ...14 Fig 6 Draft 12058 Signal and control ...
Page 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Page 18: ...18 Fig 9 Oxygen flow schematically ...
Page 19: ...19 Fig 10 Draft 12059 signal and control board ...
Page 21: ...21 Fig 12 Draft VG 96 connector ...
Page 22: ...22 Fig 13 Draft 1 14 4 1 ...
Page 23: ...23 Fig 14 Draft IR 45 power supply ...
Page 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Page 80: ...80 Abb 46 IRC 1 3 ...
Page 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Page 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Page 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Page 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Page 110: ...Service 110 ...
Page 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Page 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Page 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Page 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Page 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Page 134: ...Miscellaneous 134 Fig 47 12055 I ...
Page 135: ...Miscellaneous 135 Fig 48 12055 II ...
Page 136: ...Miscellaneous 136 Fig 49 12072 ...
Page 137: ...Miscellaneous 137 Fig 50 12064 ...
Page 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Page 139: ...Miscellaneous 139 Fig 52 12015 I ...
Page 140: ...Miscellaneous 140 Fig 53 12015 II ...
Page 141: ...Miscellaneous 141 Fig 54 12016 ...
Page 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Page 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Page 145: ...Miscellaneous 145 ...
Page 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Page 155: ......
Page 156: ......
Page 157: ......