Abb. 42:
Oscillator
Blow the dust away with compressed air. Blow from the inside to the outside,
otherwise the dust will be blown into the oscillator. Take care of the rotary speed of
the blower to remain within the operation limits. When blowing compressed air, the
rotary speed of the blower can exceed the limit of operation causing damage to the
bearings and to the shaft.
Pos : 5. 57 /0020 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Anal ysis c ycle st ops befor e t he peak begins ELTR A @ 6\ mod_1368708916668_9.doc x @ 41297 @ 233 @ 1
3.26
39B
Analysis cycle stops before the peak begins
Pos : 5. 58 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-800_Ser vic e/ 0015 F ehl er/ 019545 M odul anal ysis c ycl e st ops befor e t he peak begins ELTRA @ 6\ mod_1377688613716_9.doc x @ 46819 @ @ 1
Reasons:
The minimum analysis time is set too short in the configuration window, so that the
analysis time expires before the peak starts. Normally the peak starts about 6 to 8
seconds after the start of combustion. If the minimum analysis time is set for
example to 4 seconds, the analysis may be terminated before the peak starts. If
the minimum analysis time is set far over 8 seconds, the peak starts before the
minimum time expires. Due to the IR signal being then higher than the comparator
level, the analysis continues until the IR signal comes down to the comparator
level, unless the maximum analysis time expires earlier. If so, it is advisable to
increase the maximum analysis time in order to integrate the whole peak.
If after prolongation of the minimum analysis time to a reasonable value, a peak
still doesn't appear, possible reasons for the fault are:
–
No combustion.
–
There is no gas flow.
–
There is gas flow, but bypassing the furnace, so that there is gas flow shown
on the flow meter and by the flow sensor on the screen but no gas flow thru
the furnace. In this case the bypass valve V4 jams, keeping the oxygen flow
bypassing the furnace like in case of open furnace. The furnace has though
oxygen pressure but without having oxygen flow. Due to oxygen being
present, the metal sample and accelerator(s) burn, because there is
Summary of Contents for CS-800
Page 3: ...3 ...
Page 10: ...10 Fig 2 Draft 12057 ...
Page 11: ...11 Fig 3 Draft 11660 ...
Page 12: ...12 Fig 4 Draft 12200 ...
Page 14: ...14 Fig 6 Draft 12058 Signal and control ...
Page 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Page 18: ...18 Fig 9 Oxygen flow schematically ...
Page 19: ...19 Fig 10 Draft 12059 signal and control board ...
Page 21: ...21 Fig 12 Draft VG 96 connector ...
Page 22: ...22 Fig 13 Draft 1 14 4 1 ...
Page 23: ...23 Fig 14 Draft IR 45 power supply ...
Page 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Page 80: ...80 Abb 46 IRC 1 3 ...
Page 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Page 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Page 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Page 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Page 110: ...Service 110 ...
Page 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Page 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Page 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Page 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Page 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Page 134: ...Miscellaneous 134 Fig 47 12055 I ...
Page 135: ...Miscellaneous 135 Fig 48 12055 II ...
Page 136: ...Miscellaneous 136 Fig 49 12072 ...
Page 137: ...Miscellaneous 137 Fig 50 12064 ...
Page 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Page 139: ...Miscellaneous 139 Fig 52 12015 I ...
Page 140: ...Miscellaneous 140 Fig 53 12015 II ...
Page 141: ...Miscellaneous 141 Fig 54 12016 ...
Page 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Page 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Page 145: ...Miscellaneous 145 ...
Page 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Page 155: ......
Page 156: ......
Page 157: ......