![ELTRA CS-800 Manual Download Page 42](http://html.mh-extra.com/html/eltra/cs-800/cs-800_manual_2397410042.webp)
42
3.4.7.6 Loud buzzing sound when the generator turns on
If during integration (combustion period) the high voltage transformer is making
exceptionally loud buzzing sound when the generator turns on, also when running
the furnace empty still the loud buzzing appears, this means that obviously for
some reason, the triggering of the triac has changed.
If you are sure that the noise appears only during integration period (combustion
period) then the problem could be in the generator. The unusually loud noise could
be due to current limitation by phase angle control.
The oscillator tube may not be oscillating due to a tube fault so that it takes a very
high DC current causing the current limit circuit to reduce the phase angle. Replace
the tube. This is also the easiest thing to try so that it can be done by the customer.
NOTICE
The oscillator tube can be damaged due to overheating if for some other reason
the new tube may not be able to oscillate. For testing after installing the new tube,
put crucible with sample and accelerators into the furnace like you use to do with
every normal analysis. After starting analysis watch carefully the furnace to see the
light of the burning sample/accelerators. If a couple of seconds after generator start
(transformer noise, base lines for peak on the PC screen) there is no combustion
to see in the furnace then STOP the generator by clicking STOP on the screen.
NOTICE
When testing for induction power, always test with closed furnace.
NOTICE
When looking to find out whether there is combustion or not, STOP the analysis
cycle about 3 seconds after it started. If there are oscillations, after 3 seconds the
sample (steel or iron chips) will have visible signs of starting burning. But if you wait
up to the end of the analysis cycle in order to check whether there was
combustion, it could be already too late for the tube due to overheating.
When there are no oscillations, the tube takes a high DC current, so that after
several seconds, the anode of the tube starts glowing from the lower end. This can
be visible when the environment is not bright or it is dark.
It is practically not possible to do any measurements on the generator due to very
high voltage and due to high frequency in the oscillator (in case of functioning). So
we have to solve the problems by looking and trying.
One reason could be the resistor (8) having a short circuit. Normally its resistance
has the value of 200K.
One more possibility is a faulty capacitor. Check the capacitors, mainly the one on
top (7) of the oscillator circuit which is connecting the anode of the tube with the
oscillator circuit. Check also the grid capacitor (6) which is placed underneath the
oscillator chassis. Try to find out whether they are broken by applying a reasonable
force on them to see whether they are still in one piece. If the analyzer is relatively
old and these capacitors are in silicon sleeves, the chance that one of them is
faulty, is relatively high. If these capacitors are in brown plastic, it is less likely that
they are faulty, according to our experience.
Check also the other, bigger capacitors (9), (10) and (11) however if one of them
was faulty, most probably the fuses would blow.
Pos : 5. 13 /0020 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Bad c ombus tion ELTR A @ 6\ mod_1368608656440_9.doc x @ 40248 @ 22 @ 1
3.5
18B
Bad combustion
Pos : 5. 14 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-800_Ser vic e/ 0015 F ehl er/ 0135 Modul Schl ec hte Verbr ennung - Bad combusti on ELTRA @ 6\ mod_1377688600953_9. doc x @ 46555 @ 2 @ 1
In most cases the reasons for bad combustion are:
–
Problems with the flow
–
The Generator (oscillator) tube is not powerful enough anymore. Normally,
this does not occur suddenly. It is rather a long process.
(Part number 77210)
Summary of Contents for CS-800
Page 3: ...3 ...
Page 10: ...10 Fig 2 Draft 12057 ...
Page 11: ...11 Fig 3 Draft 11660 ...
Page 12: ...12 Fig 4 Draft 12200 ...
Page 14: ...14 Fig 6 Draft 12058 Signal and control ...
Page 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Page 18: ...18 Fig 9 Oxygen flow schematically ...
Page 19: ...19 Fig 10 Draft 12059 signal and control board ...
Page 21: ...21 Fig 12 Draft VG 96 connector ...
Page 22: ...22 Fig 13 Draft 1 14 4 1 ...
Page 23: ...23 Fig 14 Draft IR 45 power supply ...
Page 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Page 80: ...80 Abb 46 IRC 1 3 ...
Page 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Page 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Page 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Page 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Page 110: ...Service 110 ...
Page 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Page 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Page 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Page 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Page 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Page 134: ...Miscellaneous 134 Fig 47 12055 I ...
Page 135: ...Miscellaneous 135 Fig 48 12055 II ...
Page 136: ...Miscellaneous 136 Fig 49 12072 ...
Page 137: ...Miscellaneous 137 Fig 50 12064 ...
Page 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Page 139: ...Miscellaneous 139 Fig 52 12015 I ...
Page 140: ...Miscellaneous 140 Fig 53 12015 II ...
Page 141: ...Miscellaneous 141 Fig 54 12016 ...
Page 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Page 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Page 145: ...Miscellaneous 145 ...
Page 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Page 155: ......
Page 156: ......
Page 157: ......