34
Fig. 9: oxygen gauge
If the pressure is remarkably lower than 1.5 bar (22 psi) then
–
Check the pressure of the oxygen bottle.
–
Check the pressure regulator of the oxygen bottle
–
Do leakage check.
See chapter Leak Checking
3.4.2
85B
No flow
If this message permanently remains in the “status window”, check the flow on the
lower flow meter (Rotameter). The ball has to be in the range of 180 L/h. If it is in
this range, then click on the “device state” window icon.
Fig. 10: device state window icon
Check the graph for L/h. (See software helps). If the vertical bar is not in the middle
of the base line then calibrate the flow in the software as follows.
3.4.3
86B
Pre-purging
If this message permanently remains in the “status window”, click the icon of the
“configuration window”.
Fig. 11: icon configuration window
The duration of pre-purging is possibly entered much too long. Reasonable values
are between 1 and 5 seconds, maximum 10 sec for very low carbon analysis.
3.4.4
87B
Waiting for stability
If this message permanently remains in the “status window”, the infrared cell base
lines are not stable or out of range.Click the Base line Icon and check.
Summary of Contents for CS-800
Page 3: ...3 ...
Page 10: ...10 Fig 2 Draft 12057 ...
Page 11: ...11 Fig 3 Draft 11660 ...
Page 12: ...12 Fig 4 Draft 12200 ...
Page 14: ...14 Fig 6 Draft 12058 Signal and control ...
Page 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Page 18: ...18 Fig 9 Oxygen flow schematically ...
Page 19: ...19 Fig 10 Draft 12059 signal and control board ...
Page 21: ...21 Fig 12 Draft VG 96 connector ...
Page 22: ...22 Fig 13 Draft 1 14 4 1 ...
Page 23: ...23 Fig 14 Draft IR 45 power supply ...
Page 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Page 80: ...80 Abb 46 IRC 1 3 ...
Page 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Page 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Page 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Page 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Page 110: ...Service 110 ...
Page 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Page 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Page 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Page 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Page 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Page 134: ...Miscellaneous 134 Fig 47 12055 I ...
Page 135: ...Miscellaneous 135 Fig 48 12055 II ...
Page 136: ...Miscellaneous 136 Fig 49 12072 ...
Page 137: ...Miscellaneous 137 Fig 50 12064 ...
Page 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Page 139: ...Miscellaneous 139 Fig 52 12015 I ...
Page 140: ...Miscellaneous 140 Fig 53 12015 II ...
Page 141: ...Miscellaneous 141 Fig 54 12016 ...
Page 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Page 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Page 145: ...Miscellaneous 145 ...
Page 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Page 155: ......
Page 156: ......
Page 157: ......