60
–
Check if the pressure gauge for compressed air (5) shows 4 to 6 bar
(60 to 90 psi).
–
Check if the cock (3) is properly shut, or else the O-ring (4) will not seal.
Check if the O-ring (4) is in a good condition.
–
Check if the combustion tube is broken or cracked.
–
Check for gas leakage.
–
Check if the chemicals are saturated and have become hard.
–
The paper filter could also be the cause. For changing the paper filter, read
the operating manual.
3.17.2
94B
On the rear-side of the analyzer:
The contents of the gas cleaning furnace could be worn out and clogged.
3.17.3
95B
Inside the analyzer:
3.17.3.1 Oxygen solenoid valve (V5) may be defective
•
Open the left-side panel of the analyzer.
•
Unscrew the upper tube (1). The oxygen should rush out when the power
switch is set to pos. 2.
If not:
–
There may be no oxygen supplied.
Fig. 35: terminal of the solenoid coil
–
220V AC is not measured on the terminal (2) of the solenoid coil.
–
The solenoid valve is clogged.
•
Clean the solenoid valve.
If the fault is still present, then:
•
Switch off the power, (set the power switch to position 0)
•
Pull out the socket (2) from the coil and measure the coil resistance. 3.3K
is normal.
Summary of Contents for CS-800
Page 3: ...3 ...
Page 10: ...10 Fig 2 Draft 12057 ...
Page 11: ...11 Fig 3 Draft 11660 ...
Page 12: ...12 Fig 4 Draft 12200 ...
Page 14: ...14 Fig 6 Draft 12058 Signal and control ...
Page 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Page 18: ...18 Fig 9 Oxygen flow schematically ...
Page 19: ...19 Fig 10 Draft 12059 signal and control board ...
Page 21: ...21 Fig 12 Draft VG 96 connector ...
Page 22: ...22 Fig 13 Draft 1 14 4 1 ...
Page 23: ...23 Fig 14 Draft IR 45 power supply ...
Page 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Page 80: ...80 Abb 46 IRC 1 3 ...
Page 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Page 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Page 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Page 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Page 110: ...Service 110 ...
Page 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Page 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Page 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Page 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Page 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Page 134: ...Miscellaneous 134 Fig 47 12055 I ...
Page 135: ...Miscellaneous 135 Fig 48 12055 II ...
Page 136: ...Miscellaneous 136 Fig 49 12072 ...
Page 137: ...Miscellaneous 137 Fig 50 12064 ...
Page 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Page 139: ...Miscellaneous 139 Fig 52 12015 I ...
Page 140: ...Miscellaneous 140 Fig 53 12015 II ...
Page 141: ...Miscellaneous 141 Fig 54 12016 ...
Page 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Page 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Page 145: ...Miscellaneous 145 ...
Page 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Page 155: ......
Page 156: ......
Page 157: ......