![ELTRA CS-800 Manual Download Page 70](http://html.mh-extra.com/html/eltra/cs-800/cs-800_manual_2397410070.webp)
70
availability of oxygen in the furnace, coming thru the chamber valve V3
and/or thru the lance valve V2. In case of metal analysis, the burned sample
in the crucible looks like normally combusted sample. If so, check the bypass
valve V4.
–
The IR call is faulty.
Pos : 5. 59 /0020 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 C hopper mot or does n‘t r ot at e ELTR A @ 6\ mod_1368709089073_9. doc x @ 41319 @ 31221 @ 1
3.27
40B
Chopper motor doesn‘t rotate
Pos : 5. 60 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-800_Ser vic e/ 0015 F ehl er/ 019550 M odul C hopper mot or does n‘t r ot at e ELTR A @ 6\ mod_1377688613914_9. doc x @ 46831 @ 2 @ 1
The rotation of the chopper motor needs to be checked in case of problems only,
like when there are no combustion peaks on the screen or the IR signals are minus
-10V (take care, minus) displayed on the PC screen in the window “inputs”.
In order to check whether the chopper motor is rotating, the easiest way is to
remove an IR path.
See chapter
IR-Paths cleaning and replacing
.
Mostly, the easiest path to remove is the shortest one. Then, the chopper blade
can be seen thru the hole of the chopper housing, which becomes free (open) after
removing the path.
The reason of failure can be a faulty motor but also the controlling electronics of
the motor may have become defective. If so, the motor or the IRC board has to be
replaced. Before replacing, check at least the voltages on the board, mainly the
supply voltage of 24VDC.
See chapter
Infrared base line adjustments
.
For replacing the chopper see chapter
Chopper
.
Pos : 5. 61 /0020 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 Sample burns but no peak on the s creen @ 7\ mod_1381400077047_9. doc x @ 49828 @ 222 @ 1
3.28
41B
Sample burns but no peak on the screen
Pos : 5. 62 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-800_Ser vic e/ 0015 F ehl er/ 019555 M odul Sample burns but no peak on screen @ 7\ mod_1381400241865_9. doc x @ 49839 @ 3 @ 1
There is gas flow, but bypassing the furnace, so that there is gas flow shown on
the flow meter and by the flow sensor on the screen but no gas flow thru the
furnace.
In this case the bypass valve V4 jams, causing the oxygen flow to bypass the
furnace like in case of open furnace.
See chapter “
Gas flow regulation
.
The furnace has though oxygen pressure but without having oxygen flow. Due to
oxygen being present, the metal sample and accelerator(s) burn, because there is
availability of oxygen in the furnace, coming thru the chamber valve V3 and/or thru
the lance valve V2. In case of metal analysis, the burned sample in the crucible
looks like normally combusted sample. If so, check the bypass valve V4.
Pos : 5. 63 /0020 Übersc hrift en/ELTRA 1.1 Übersc hrift en/11 IR signal does not c ome down t o the base line at t he end of anal ysis @ 7\ mod_1381404509510_9. doc x @ 49850 @ 31222 @ 1
3.29
42B
IR signal does not come down to the base line at the end of analysis
Pos : 5. 64 /0010 ELTR A/ 0015 s ervic e_ins truc tions/C S-800_Ser vic e/ 0015 F ehl er/ 019560 M odul IR signal does not c ome down t o the base line at the end of anal ysis @ 7\ mod_1381404613105_9. doc x @ 49862 @ 3323333 @ 1
There are several different reasons for this effect:
Summary of Contents for CS-800
Page 3: ...3 ...
Page 10: ...10 Fig 2 Draft 12057 ...
Page 11: ...11 Fig 3 Draft 11660 ...
Page 12: ...12 Fig 4 Draft 12200 ...
Page 14: ...14 Fig 6 Draft 12058 Signal and control ...
Page 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Page 18: ...18 Fig 9 Oxygen flow schematically ...
Page 19: ...19 Fig 10 Draft 12059 signal and control board ...
Page 21: ...21 Fig 12 Draft VG 96 connector ...
Page 22: ...22 Fig 13 Draft 1 14 4 1 ...
Page 23: ...23 Fig 14 Draft IR 45 power supply ...
Page 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Page 80: ...80 Abb 46 IRC 1 3 ...
Page 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Page 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Page 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Page 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Page 110: ...Service 110 ...
Page 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Page 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Page 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Page 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Page 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Page 134: ...Miscellaneous 134 Fig 47 12055 I ...
Page 135: ...Miscellaneous 135 Fig 48 12055 II ...
Page 136: ...Miscellaneous 136 Fig 49 12072 ...
Page 137: ...Miscellaneous 137 Fig 50 12064 ...
Page 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Page 139: ...Miscellaneous 139 Fig 52 12015 I ...
Page 140: ...Miscellaneous 140 Fig 53 12015 II ...
Page 141: ...Miscellaneous 141 Fig 54 12016 ...
Page 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Page 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Page 145: ...Miscellaneous 145 ...
Page 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Page 155: ......
Page 156: ......
Page 157: ......