The temperature is set by turning the pot and measuring the set value on
The actual value of the temperature is measured on TP26.
Caution: After changing the temperature set point, it takes some time until the
actual temperature value is stabilized at the new set level.
This is the trimmer in the IR cell module for adjusting the cell temperature. It is
accessible with a screw driver from outside the cell. There is no need to remove
the cell block module from the analyzer’s cabinet in order to adjust the
temperature.
4.2.8.2 Modifications for higher sensitivity (gain) in the low sulphur range
For LS: R101=47k R100 = 3M
4.1 x Standard gain
Recalculation factor = 92.664
For double the normal sensitivity: R101=180k
1.92 x standard gain
Recalculation factor = 39.5
For standard gain.
Recalculation factor = 20.59
4.2.8.3 IRC board and cross-talk
For non-soldered sections, at place of the capacitors C90 to C93, jumpers must be
soldered (short pieces of wire) to connect not used IR outputs to GND.
There is a peak to see on the screen if not available IR cells are selected in the PC
software. For example if an analyzer has one carbon cell only, but the operator
selects the configuration of one C and one S ranges in the PC software, he will see
a sulfur peak on the screen giving also sulfur results although there is no sulfur cell
existing in the IR rack. The reason is cross-talk in the signal cable and in the
multiplexer also due to the high input impedance of the UNI board.
In order to avoid this, the outputs of not existing IR ranges have to be connected to
GND on the IRC board, so that there will be no peak on the corresponding UNI
input and on the PC screen.
Summary of Contents for CS-800
Page 3: ...3 ...
Page 10: ...10 Fig 2 Draft 12057 ...
Page 11: ...11 Fig 3 Draft 11660 ...
Page 12: ...12 Fig 4 Draft 12200 ...
Page 14: ...14 Fig 6 Draft 12058 Signal and control ...
Page 17: ...17 Fig 8 Draft 6 2 4 Gas flow diagram furnace closed ...
Page 18: ...18 Fig 9 Oxygen flow schematically ...
Page 19: ...19 Fig 10 Draft 12059 signal and control board ...
Page 21: ...21 Fig 12 Draft VG 96 connector ...
Page 22: ...22 Fig 13 Draft 1 14 4 1 ...
Page 23: ...23 Fig 14 Draft IR 45 power supply ...
Page 24: ...24 Fig 15 Draft IR sources Fig 16 Draft 7 pin plug ...
Page 80: ...80 Abb 46 IRC 1 3 ...
Page 82: ...82 Fig 29 Mainboard Cut out IR Cell general test points ...
Page 91: ...91 Abb 47 Pneumatics parts 2 1 3 4 5 6 7 8 9 10 11 12 13 14 15 ...
Page 94: ...94 bar 0 5 1 1 5 2 2 5 0 PR2 1 Fig 49 Front side 4 5 47BPurge pressure Fig 50 PR2 5 6 ...
Page 109: ...Service 109 10 10 11 9 b7 10 1 12 8 5 6 Fig 60 b7 10 1 ...
Page 110: ...Service 110 ...
Page 119: ...Miscellaneous 119 77510 Heaters for IR cell ...
Page 120: ...Miscellaneous 120 6 1 2 121BFront side Fig 41 Front side ...
Page 122: ...Miscellaneous 122 6 1 3 122BRight side Fig 42 Right side ...
Page 126: ...Miscellaneous 126 6 1 5 124BOscillating circuit Fig 44 Oscillating circuit ...
Page 132: ...Miscellaneous 132 21010 Heather section 21120 Quartz tube 70380 O Ring 70380 21120 21010 ...
Page 134: ...Miscellaneous 134 Fig 47 12055 I ...
Page 135: ...Miscellaneous 135 Fig 48 12055 II ...
Page 136: ...Miscellaneous 136 Fig 49 12072 ...
Page 137: ...Miscellaneous 137 Fig 50 12064 ...
Page 138: ...Miscellaneous 138 Fig 51 12550 Vacuum cleaner ...
Page 139: ...Miscellaneous 139 Fig 52 12015 I ...
Page 140: ...Miscellaneous 140 Fig 53 12015 II ...
Page 141: ...Miscellaneous 141 Fig 54 12016 ...
Page 143: ...Miscellaneous 143 6 2 1 127BUNI 1 ...
Page 144: ...Miscellaneous 144 6 2 2 128BGas flow system ...
Page 145: ...Miscellaneous 145 ...
Page 146: ...Miscellaneous 146 6 2 3 129BValve controlling Fig 55 Valve controlling I 11001 6001 ...
Page 155: ......
Page 156: ......
Page 157: ......