Elma Schmidbauer GmbH
About this manual |
Elmasonic Med
31 / 32
ü
The tank is empty.
1. Use a damp cloth to wipe out the remaining limescale and dirt residue. If necessary, rinse the
tank using a shower set.
CAUTION! Do not spray the outside of the unit with water.
ð
The tank has been cleaned.
2. To descale the tank, fill it with tap water up to the fill level marking in the tank.
3. Heat the unit up to 40 °C by pressing the
Heating
button.
4. Add 4–10% of Elma clean 60 or Elma clean 115C.
5. Once 40 °C have been reached, switch off the unit.
ð
Leave the solution to work for approx. 12 hours.
6. Then, switch on
Eco
ultrasound mode for approx. 15 min.
7. Empty the tank.
8. Use a damp cloth to wipe out the remaining limescale and dirt residue. If necessary, rinse the
tank using a shower set.
CAUTION! Do not spray the outside of the unit with water.
è
The tank has been descaled and cleaned.
Disinfect:
We recommend the following disinfectants for wiping the unit:
• Incidin Liquid by Ecolab (ready-to-use, quick disinfectant).
• Terralin protect by Schülke (observe dosage and exposure time).
ü
The unit is used in the medical and health sector.
NOTICE! Observe local directives and cleaning regulations. This particularly applies to the
cleaning of medical devices.
◦ Disinfect the tank and the enclosure using a commercially available surface disinfectant on a
regular basis.
CAUTION! Test the disinfectant on a small area first to ensure it does not
affect the materials, particularly the controls.
è
The unit has been hygienically disinfected.
Clean the blower, and check it is working:
Dust and dirt adhering to the protective grille of the blower can impair the cooling performance or
cause unit malfunctions.
• Regularly remove accumulated dust and dirt.
ü
The unit is ready for operation and filled with fluid to enable the functions to be checked.
1. Heat the unit up to approx. 55 °C by pressing the
Heating
button.
2. The blower will start after > 50 °C is reached.
ð
If the blower is defective, the sound made by it will be clearly audible or the display will show
the error message Er 47.
3. Press the
On/Off
buttons to switch off the unit.
The blower function has been checked. If no sound can be heard from the blower or the error
message is shown, send the unit for repair immediately.
Test LED function:
Regularly check the function of all LEDs on the display to ensure reliable operation. The display
showing all LED lamps is illustrated in the chapter
Controls
. Send the unit for repair immediately if any
LED lamps stop working.
The unit is switched off.