Elma Schmidbauer GmbH
About this manual |
Elmasonic Med
27 / 32
ü
The basket or insert has been removed.
ü
If using temperatures > 50 °C, allow the cleaning fluid to cool.
1. Disconnect the network cable.
2. Put the lid on.
3. Take hold of the unit firmly by the handles holding the lid in place, and carry it carefully to a sink.
4. Remove the lid.
5. Slowly tip the unit over the sink with one corner at the lowest point, and carefully pour out the
cleaning fluid.
ð
The tank is empty.
6. Clean the tank, and disinfect it if necessary. See chapter
Maintenance
for information.
CAUTION! Never submerge the unit in water.
è
The unit is emptied, cleaned and, where necessary, disinfected.
14
Change parameters in Setup
IMPORTA
NT
If parameters marked with a star (*) are changed, the saved programme will be deleted.
If necessary, make a note of the settings of the saved programmes before changing
them.
The device is connected to a power supply.
The
Parameter reference
list is required to set the parameter numbers and values.
The unit is switched off.
Hold down the
Mode
button until the
Setup
menu is shown on the display.
First, press the
Plus/Minus
buttons to set the parameter number, e.g. 01.
Then, press the
Plus/Minus
buttons to change the parameter value, e.g. 00/01.
Set and change further parameters as required.
Press the
On/Off
button to save the settings.
Parameter reference:
The following unit functions can be adjusted. The preset factory values are shown in bold:
Parameter
Unit function
Value
Effect