Elma Schmidbauer GmbH
About this manual |
Elmasonic Med
23 / 32
Set cleaning duration
(1 min. to 6 h):
Press one of the
Plus/Minus
buttons below the time display to set the duration.
Press the button repeatedly until the required duration is shown on the display. Hold
the button down to increase/decrease the duration more quickly.
Set ultrasound mode
(
Degas
,
Eco
,
Sweep
,
Pulse
or
Dyn
):
Press the
Mode
button to set the ultrasound mode. Press the button repeatedly until
the required ultrasound mode is shown on the display.
Set target temperature
(25–80°C/77–176°F):
Press one of the
Plus/Minus
buttons below the temperature display to set the
required target temperature. Press the button repeatedly until the required
temperature flashes on the display. Hold the button down to increase/decrease the
duration more quickly.
IMPORTANT
If a temperature limit between 40–60°C/104–140°F has been set, see
the section
Change parameters
for information on changing the temperature limit.
Longer periods of ultrasound operation heat the cleaning fluid in addition to the
heating function. This can lead to the fluid being hotter than the set target
temperature.
The
Auto-degas
function can also be set to degas a fresh batch of cleaning fluid.
Set Auto-degas:
One of the ultrasound modes
Eco
,
Sweep
,
Pulse
or
Dyn
is set.
Ensure that the unit is not in ultrasound operation.
Hold down the
Start/Stop
button until the ultrasound mode
Degas
flashes on the
display.
Press the
Start/Stop
button again to start the
Auto-degas
function.
Alternatively, save
Auto-degas
as a programme. See the section
Save programmes
for more information.
Degas
flashes slowly on the display (degassing takes approx. 10 min.). Once
Auto-
degas
has been completed, the set ultrasonic cleaning mode will start and be
shown on the display.
Set the
Temperature-controlled start
function to start ultrasound operation automatically once the
target temperature has been reached.
IMPORTANT
When cleaning sensitive objects, please note that the ultrasound is switched on briefly
multiple times during the heating phase.
Set Temperature-controlled start:
Ensure that the required target temperature is set.