6.4.22 Condenser probe (black)
Steps for replacing the condenser probe:
154.
Follow the procedure in the [
Upper Panel
] and [
Electrical Bo
x] sections to reach the probe con-
nector and disconnect it.
155.
Pull out probe from its position (Fig. 1).
156.
Insert the bulb of the new probe into the seat.
157.
Use cable ties to restore the wiring layout.
158.
Reassemble all the removed components following
the inverse procedure
.
6.4.23 Tanic Receiver
Steps for replacing the Tanic receiver:
ATTENTION!
Before replacing the Receiver, make sure that the suction valve and the exhaust gas valve are closed and follow
the important procedures shown on page 103.
159.
Remove the front panel as explained in the paragraph [
Upper Panel
].
160.
Desolder the two connection pipes to the receiver, as shown in figure 1.
161.
Turn the receiver counterclockwise to remove it from its seat (Fig. 2).