D13510840D - RC Remote control
Page 26
Characteristics
Environmental Conditions
Mains power for wall plug-in power supply
Wireless permits of ELD500 RC Wireless
CE, FCC, IC, TELEC, MIC, MII
Audio alarm at 1 m distance
70 db(A) maximum
Headphone jack stereo 3.5 mm
> 2 x 32 Ohm
Range of wireless transmission
> 100 m in free field
HF output power
+6 dBm (4 mW)
Frequency of wireless transmission
2.4 GHz
Internal memory capacity
64 MB, 32 MB of which is available for
recording data
Battery operation time (ELD500 RC Wireless)
> 8 hours (depending on battery level)
Display
TFT-Touch 1/4 VGA / 3.5 inch
240 x 320 px, maximum error 4 px
ELD500 RC Wireless charging voltage from power
supply
24 V d.c., maximum 0.7 A
ELD500 RC Wired supply voltage from leak
detector
24 V d.c., maximum 0.7 A
Only for use within buildings.
EN 61010
Permissible ambient temperature (during
operation)
+5 to +40 °C
Permissible storage temperature
-10 to +60 °C
Maximum relative humidity
80% to 31 °C, linear decreasing 50% at
40 °C
Maximum permissible height above sea level
(during operation)
2000 m
Mains voltages and frequencies
100 - 250 V a.c. 50/60 Hz
Interchangeable blades of mains plug
Europe, North America, Japan, UK,
China, Australia
Power consumption
maximum 30 VA
Summary of Contents for ELD500 Series
Page 9: ...D13510840D Contents Page vii...
Page 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Page 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Page 38: ...D13510840D Page 28 This page has been intentionally left blank...
Page 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Page 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Page 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Page 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Page 121: ......