D13510840D - Interface description
Page 106
Analog output leak rate scale (log. only) (command 223)
State calibration (command 260)
Value
Meaning
0
0.5 V / decade
1
1 V / decade
2
2 V / decade
3
2,5 V / decade
4
5 V / decade
5
10 V / decade
Value
Index 0: Meaning
0
Task idle
1
Task start
2
Task running
3
Task stop
Value
Index 1: Meaning
0
CAL idle
1
CAL internal (no user action required)
2
CAL external (acknowledge at "6" required)
Value
Index 2: Meaning
0
CAL_IDLE
1
CAL_EVAC
2
CAL_WAIT_STABLE_OPEN
3
CAL_TUNE
4
CAL_TUNE_RESTART
5
CAL_WAIT_CLOSE
6
CAL_WAIT_STABLE_CLOSE
7
CAL_WAIT_OK