D13510840D - Calibrated leak - internal
Page 55
Calibrated leak - internal
Description
General
These operating instructions contain important information on how to mount the test leak
CL - Internal.
Unpack the calibrated leak CL - Internal immediately after delivery, even if it is to be put into
operation at a later date.
Check the calibrated leak CL - Internal for completeness (see
) and carefully
examine it visually. If any damage is discovered, report it immediately to the forwarding
agent and insurer. If the damaged part has to be replaced, please get in touch with the orders
department.
Technical Data
Purpose
The calibrated Leak CL - Internal with the helium reservoir is used for alignment of the helium
mass spectrometer in the helium leak detector as well as for calibrating the leak rate
indication.
It is equipped with a solenoid valve which is actuated via the control electronics of the leak
detector.
Note:
Indicates special technical requirements that the user must comply with.
Note:
The shipping container and packaging materials must be kept in the event of complaints about
damage.
Nominal calibration range
10
-7
mbar l/s
Tolerance of nominal calibration range
± 15%
Temperature coefficient
< 0.5% / °C
Leak type
Capillary
Calibrated for
Helium
Filling
Helium
Filling pressure
2.9 bar
Connecting flange
Special adaptor
Summary of Contents for ELD500 Series
Page 9: ...D13510840D Contents Page vii...
Page 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Page 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Page 38: ...D13510840D Page 28 This page has been intentionally left blank...
Page 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Page 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Page 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Page 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Page 121: ......