In the laminar flow region it applies:
= dynamic viscosity
Some values of dynamic viscosity
in 10
-5
Pa s.
The resulting values must be used according to the predominant type of gas flow, i.e. for leak
rates lower than 10
-5
mbar Is
-1
mainly molecular flow and for leak rates higher than
10
-5
mbar Is
-1
mainly laminar flow can be assumed.
Adjustment of leak rate
The nominal leak rate was measured during production at a test gas pressure of 1 bara versus
a pressure of lower than 1 mbar at the connecting flange and is shown on the label. The
adjustment diagrams are valid for nominal leak rate of 5 x 10
-6
mbar ls ±10% for the CL - He
4-6. The leak rate is pressure dependent and can be found in
or
. Upon request,
individual calibration of a single test leak as a function of the search-gas pressure can be made
in our factory at cost price.
If you suspect the nominal leak rate has changed after longer use, recalibration of the test
leak can be made in our factory against charge. When demanding higher accuracy Edwards
recommends a recalibration once a year.
If a calibration certificate is required, it can be issued by Service. Calibration and calibration
certificates are available to order, charges applicable.
The test leak rates depend on the adjusted pressure [bar rel], see
.
If a significant deviation from the nominal leak rate of 5 x 10
-6
mbar ls occurs, a factor can be
calculated based on the leak rate diagrams. The nominal leak rate is to be multiplied by this
factor.
Calculation of the factor:
He
R 134a
N2O2
Ar
H2
h
1.96
1.36
1.82
2.21
0.88
q
x
q
He
--------
He
x
---------
=
Summary of Contents for ELD500 Series
Page 9: ...D13510840D Contents Page vii...
Page 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Page 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Page 38: ...D13510840D Page 28 This page has been intentionally left blank...
Page 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Page 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Page 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Page 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Page 121: ......