D13510840D - Partial flow adaptor
Page 51
Examples:
For a quick point time of T
Q
= 0 seconds and by operating "START", the valve V10 is initially
not opened. This setting is recommended in the case of large volumes and dirty test samples.
For T
Q
= ∞ (>>) valve V10 is opened upon operating "START". At an inlet pressure of pE < 3
mbar the leak detector will enter into the measurement mode and display leak rates. V10
remains open until pressing "STOP". The setting of T
Q
is recommended in such cases where
it is acceptable to wait until the leak detector enters into the measurement mode and when
it is not necessary to display a leak rate before that time.
Equipment
Standard specification
Partial flow valve block
DN 25 KF right angle valve
Control cable
Vacuum hose
Small flange connections
90° elbow/pipe bend
Operation
For proper start up and operation of the partial flow adaptor it is absolutely necessary to take
note of the information provided in the separate operating instructions for the components
of the system besides the information given in these operating instructions.
Start up
An overview of the arrangement of the partial flow adaptor is given in
Switch the ELD500 off and disconnect any possibly connected test objects from the test port.
Connect the partial flow valve block (see
item 2) to the test port of the ELD500.
When doing so, the metal filter on the rear should point backwards and the lateral vacuum
flange should point to the left.
In order to exclude the effects of any interference, the right-angle valve DN 25 KF
item 1) must now be flanged to the side port of the partial flow valve block
item 2) so that the second flange points to the front and so that the LEDs of the
valve point to the left.
The partial flow pump (see
item 8) is connected via the vacuum hose
item 9) and the pipe bend/elbow (see
item 10) to the DN 25 KF right-
item 1). Provide the electrical connection through connection cable
item 5) as follows:
1. Screw together the round connection plug with the cable coupling of the right angle
valve.
2. Connect plug (see
Summary of Contents for ELD500 Series
Page 9: ...D13510840D Contents Page vii...
Page 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Page 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Page 38: ...D13510840D Page 28 This page has been intentionally left blank...
Page 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Page 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Page 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Page 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Page 121: ......