D13510840D - Calibrated leak - He 4-6
Page 63
Changing the reservoir cylinder without changing the type of gas
1. Close all valve (the knob turns upwards).
2. Unscrew the old test gas bottle.
3. After removing the protective cap screw on the new test gas bottle.
4. Open the test gas supply. See
, item 2.
The manometer displays now the test gas bottle pressure.
5. Close the test gas supply.