D13510840D - Calibrated leak - internal
Page 56
Supplied parts
The supplied parts of the CL - Internal comprises in addition:
O-ring 12 x 1.5
Two cheese head screws M 3 x 25
Sticker stating the leak rate
Assembly
Tools required
Cross-head screwdriver, size 1
Hex socket screw key, size 2.5
Installing/uninstalling the CL - Internal
See also documentation of the leak detector.
Loosen CL - Internal from the valve block by unscrewing the screws M3. Take off the electrical
connection. For installing take care that the o-ring and the sealing surfaces are clean.
Leak tightness test
Start the helium leak detector and spray a little helium through the intake opening of the TMP
fan into the leak detector.
Entering the value of the calibrated leak
To enter the value of the calibrated leak please refer to the Technical Handbook, Section "The
menu function (Overview)" or the "Description of the individual menu functions" under menu
item "Cal leak 0.0E-07".
Factory inspection
Calibrated leaks are not subject to wear and the helium loss of the calibrated leak
CL - Internal, being less than 2% per year, is negligible. Nevertheless, the leak rate may change
over years due to external influences. A factory inspection is, therefore, advisable once a
year.
A test certificate, if required for the helium calibrated leak, can be obtained through Edwards.
In that case, the calibrated leak should be forwarded to us and will be returned inspected and
recertified with the test certificate, charges applicable.
The helium flow rate stated on the main label is the actual leak rate of the calibrated leak. See
Note:
In the range of 10
-9
mbar l/s no indication must appear.
Summary of Contents for ELD500 Series
Page 9: ...D13510840D Contents Page vii...
Page 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Page 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Page 38: ...D13510840D Page 28 This page has been intentionally left blank...
Page 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Page 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Page 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Page 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Page 121: ......